Installation: Preparation for Use
2.
Press the Internal Access Button to open the MAC 5500, then plug the cable
connector into port
A.
Opening the MAC 5500 before attaching the cable clamp
allows you to place the correct amount of slack to free the cable from stress
when the MAC 5500 needs to be re-opened.
3.
Next fasten the cable and clamp to the clamp bracket, then close the MAC 5500.
Observe that there is enough slack to allow free movement of the cable when re-
opening the MAC 5500.
4.
Refer to the MAC 5500 Operator’s Manual for information on how to use the
Bar Code Reader.
Port A
Internal Access
Button
74A
Not enough cable slack.
Correct amount of cable slack.
Summary of Contents for MAC 5500
Page 1: ...g MAC 5500 Resting ECG Analysis System Service Manual 2020299 020 Revision E GE Healthcare ...
Page 9: ...Revision E MAC 5500 resting ECG analysis system 1 1 2020299 020 Introduction 1 Introduction ...
Page 10: ...1 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Introduction For your notes ...
Page 55: ...Revision E MAC 5500 resting ECG analysis system 3 1 2020299 020 Installation 3 Installation ...
Page 56: ...3 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Installation For your notes ...
Page 105: ...Revision E MAC 5500 resting ECG analysis system 5 1 2020299 020 Maintenance 5 Maintenance ...
Page 106: ...5 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Maintenance For your notes ...
Page 137: ...Revision E MAC 5500 resting ECG analysis system 6 1 2020299 020 Parts List 6 Parts List ...
Page 138: ...6 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Parts List For your notes ...
Page 175: ...Revision E MAC 5500 resting ECG analysis system Index 1 2020299 020 Index ...
Page 176: ...Index 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 178: ...Index Index 4 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 179: ......