Revision E
MAC
5500 resting ECG analysis system
5-17
2020299-020
Maintenance: Disassembly Guidelines
the label of the system.
6.
Select/verify that
No
is selected for
Update Primary Boot
.
7.
Enter the
Print head resistance
. This number can be found on the print head
label.
8.
Select the appropriate language in the
Keyboard
menu.
9.
Select
Return
.
Restore System Setups
1.
Power up the cart.
2.
From the
Main Menu
, select
System Setup
.
3.
Enter the
System Password
and press
Enter
.
4.
Select
Restore Setup
from the
System Setup
menu.
5.
Select
From SD Card
from the
Restore Setup
menu.
Restore Options
Using the option activation codes for the system, restore the options which had been
installed on the board which was removed. These options are printed on a label
located on the bottom of the paper tray.
NOTE
Use the activator codes shown on the label on your system. The activator codes
shown in the figure below are examples only.
1.
Power on the cart.
2.
Within the
System Setup
function, select
Basic System
.
3.
Select
Option Activation
to activate options.
4.
Type the 12-digit option activation code and press the
Enter
key.
5.
Repeat the previous step for each option to be activated on the new system.
122A
Summary of Contents for MAC 5500
Page 1: ...g MAC 5500 Resting ECG Analysis System Service Manual 2020299 020 Revision E GE Healthcare ...
Page 9: ...Revision E MAC 5500 resting ECG analysis system 1 1 2020299 020 Introduction 1 Introduction ...
Page 10: ...1 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Introduction For your notes ...
Page 55: ...Revision E MAC 5500 resting ECG analysis system 3 1 2020299 020 Installation 3 Installation ...
Page 56: ...3 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Installation For your notes ...
Page 105: ...Revision E MAC 5500 resting ECG analysis system 5 1 2020299 020 Maintenance 5 Maintenance ...
Page 106: ...5 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Maintenance For your notes ...
Page 137: ...Revision E MAC 5500 resting ECG analysis system 6 1 2020299 020 Parts List 6 Parts List ...
Page 138: ...6 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Parts List For your notes ...
Page 175: ...Revision E MAC 5500 resting ECG analysis system Index 1 2020299 020 Index ...
Page 176: ...Index 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 178: ...Index Index 4 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 179: ......