Troubleshooting: Frequently Asked Questions
Passwords
Q: Can you set up a password for the
Delete
function that is different than the
System Setup
password?
A: No. The password for the
System Setup
and the
Delete
function are the same.
Clinical
Report Format
Q: How do I change the way an ECG looks when it prints out?
A: Do the following steps.
Go to
System Setup
.
Select
ECG
.
Select which type of ECG report you want to change:
Resting ECG Reports
Pediatric ECG Reports
15 Lead Reports
Select unconfirmed reports from the menu.
Find the report type you want the MAC 5500 to print.
Place the number of copies you want in the appropriate column.
If you want the MAC 5500 or 12SL interpretation included on the ECG, put the
number of copies you want in the “with” column.
If you do not want the MAC 5500 interpretation to print on the ECG, put the
number of copies you want in the “without” column.
Click view report type, to see the examples of the report formats.
Press the
esc
key until you return to
System Setup
.
Select
Save Setup
.
Select
To System
.
You can select
Main Menu
to exit
System Setup
.
Hi-Res and Phi-Res
Q: What is the difference between the Hi-Res and Phi-Res functions?
A: Hi-Res looks at the entire complex, whereas Phi-Res focuses on the P wave.
Editing
Q: Can you edit the interpretation at the MAC 5500, and then transmit the edited
record to the MUSE system as an unconfirmed record?
A: If you edit demographic information only the record is still transmitted to the
MUSE system as an unconfirmed record. However, if you edit the
interpretation, the data will not be saved unless the record is confirmed at the
MAC 5500. The record is transmitted to the MUSE system as a confirmed
record as well.
Summary of Contents for MAC 5500
Page 1: ...g MAC 5500 Resting ECG Analysis System Service Manual 2020299 020 Revision E GE Healthcare ...
Page 9: ...Revision E MAC 5500 resting ECG analysis system 1 1 2020299 020 Introduction 1 Introduction ...
Page 10: ...1 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Introduction For your notes ...
Page 55: ...Revision E MAC 5500 resting ECG analysis system 3 1 2020299 020 Installation 3 Installation ...
Page 56: ...3 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Installation For your notes ...
Page 105: ...Revision E MAC 5500 resting ECG analysis system 5 1 2020299 020 Maintenance 5 Maintenance ...
Page 106: ...5 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Maintenance For your notes ...
Page 137: ...Revision E MAC 5500 resting ECG analysis system 6 1 2020299 020 Parts List 6 Parts List ...
Page 138: ...6 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Parts List For your notes ...
Page 175: ...Revision E MAC 5500 resting ECG analysis system Index 1 2020299 020 Index ...
Page 176: ...Index 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 178: ...Index Index 4 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 179: ......