B-4
MAC
5500 resting ECG analysis system
Revision E
2020299-020
Appendix B – Technical Specifications: Technical Specifications
Writer
Item
Description
Type
Thermal dot array
Speeds
5, 12.5, 25, 50 mm/s (same as display)
Number of Traces
3, 6, 12, or 15 user-selectable (same as display)
Sensitivity/Gain
2.5, 5, 10, 20, 10/5 (split calibration) mm/mV (same as display)
Speed Accuracy
± 2%
Amplitude Accuracy
± 5%
Resolution
Horizontal 1000 dpi @ 25 mm/s, 200 dpi vertical
Paper Type/Size
Thermal z-fold, perforated, 215.9 mm x 276.4 mm (8.5 in. x 11 in.) fanfold, 300 sheets per
pack
Keyboard
Item
Description
Type
Sealed elastomer with soft function keys, alphanumeric keys, writer controls and TrimPad
cursor controls
Electrical
Item
Description
Power Supply
AC or battery operation
AC Input
Voltage
Current
Frequency
100-240 VAC, +10, -15%
0.5 A @ 115 VAC, 0.3 A @ 240 VAC, typical
50/60 Hz, ±10%
Battery Type
User replaceable, 18V @ 3.5 AH ±10%, rechargeable NiMH pack
Battery Capacity
100 single-page reports (typical) or 6 hours continuous operation (without printing)
Battery Charge Time
Approximately 4.5 hrs. from total discharge with display off.
NOTE: Cannot charge battery at or above 45° C (best if below 40° C)
Vectorcardiography
Item
Description
Report Formats
Vector loops of component vectors (P, QRS, ST-T)
Sensitivity
20, 40, 80, 160 mm/mV
Time Resolution
2 ms
Summary of Contents for MAC 5500
Page 1: ...g MAC 5500 Resting ECG Analysis System Service Manual 2020299 020 Revision E GE Healthcare ...
Page 9: ...Revision E MAC 5500 resting ECG analysis system 1 1 2020299 020 Introduction 1 Introduction ...
Page 10: ...1 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Introduction For your notes ...
Page 55: ...Revision E MAC 5500 resting ECG analysis system 3 1 2020299 020 Installation 3 Installation ...
Page 56: ...3 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Installation For your notes ...
Page 105: ...Revision E MAC 5500 resting ECG analysis system 5 1 2020299 020 Maintenance 5 Maintenance ...
Page 106: ...5 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Maintenance For your notes ...
Page 137: ...Revision E MAC 5500 resting ECG analysis system 6 1 2020299 020 Parts List 6 Parts List ...
Page 138: ...6 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Parts List For your notes ...
Page 175: ...Revision E MAC 5500 resting ECG analysis system Index 1 2020299 020 Index ...
Page 176: ...Index 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 178: ...Index Index 4 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 179: ......