Troubleshooting: Diagnostic Tests
Battery Tests
NOTE
The minimum discharge capacity is 2000 mAH. Consider replacing the battery
if this number is less than 2000 mAH.
Battery Status
Displays, and constantly updates, the following information:
Percent of charge remaining
Battery voltage (With a reading of 80% or more for percent of charge
remaining, the battery voltage should be between 15 and 24 volts. If battery
voltage is below 15 volts, the battery may need to be replaced.)
Battery current
If battery current is less than -0.7 amps with AC power not applied, the main
CPU may need to be replaced.
(For example: Consider replacing the main CPU if the battery current is -0.8
amps with AC power not applied.)
Battery temperature.
Maximum and minimum battery temperature
Ambient temperature (inside the unit).
If the ambient temperature (inside the unit) is more than 10
greater than the
current room temperature.
Maximum and minimum ambient temperature
Current battery charging status
Battery Discharge Test
This test charges the battery to full capacity, if necessary, then monitors a discharge
cycle.
NOTE
To cancel the test at anytime, press the
ESC
key.
1.
To perform the battery discharge test, plug the unit into AC (mains) power.
2.
Select
Battery Discharge Test
. The battery discharge test window will appear
and the unit will begin to charge the battery.
3.
Once the battery is fully charged, the message
Turn the AC power to the unit
OFF!
will appear.
4.
Unplug the unit from AC (mains) power and select
OK
.
5.
Re-select
Battery Discharge Test
. The battery will begin to discharge. When the
battery has fully discharged, the unit will shut off.
6.
Reconnect the unit to AC power and turn the unit on. Go to
Print Discharge
Test Results
.
Monitored information, written to the internal memory, includes:
Discharge capacity (in mAH)
Summary of Contents for MAC 5500
Page 1: ...g MAC 5500 Resting ECG Analysis System Service Manual 2020299 020 Revision E GE Healthcare ...
Page 9: ...Revision E MAC 5500 resting ECG analysis system 1 1 2020299 020 Introduction 1 Introduction ...
Page 10: ...1 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Introduction For your notes ...
Page 55: ...Revision E MAC 5500 resting ECG analysis system 3 1 2020299 020 Installation 3 Installation ...
Page 56: ...3 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Installation For your notes ...
Page 105: ...Revision E MAC 5500 resting ECG analysis system 5 1 2020299 020 Maintenance 5 Maintenance ...
Page 106: ...5 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Maintenance For your notes ...
Page 137: ...Revision E MAC 5500 resting ECG analysis system 6 1 2020299 020 Parts List 6 Parts List ...
Page 138: ...6 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Parts List For your notes ...
Page 175: ...Revision E MAC 5500 resting ECG analysis system Index 1 2020299 020 Index ...
Page 176: ...Index 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 178: ...Index Index 4 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 179: ......