Revision E
MAC
5500 resting ECG analysis system
5-21
2020299-020
Maintenance: Disassembly Guidelines
5.
Enter the
IP Address, Subnet Mask,
and
Gateway
provided by the network
administrator
6.
Enter the
Port Number
used by the MUSE system.
7.
Select
Return
and press
Enter
.
The Network Settings window closes.
8.
On the Basic System menu, select Transmission and press Enter.
The
Transmission
window opens.
9.
Enter the following information.
10. Select
Return
and press
Enter
.
The
Transmission
window closes.
11. On the
Basic System
menu, select
Return
and press
Enter
.
The
Basic System
menu closes.
12. On the
System Setup
menu, select
Save Setup
>
To System
.
This saves the settings to the communication board. This should take
approximately 8 to 10 seconds. If it takes less than 8 seconds, the settings were
not saved successfully. Repeat a second time.
13. Cycle power on the MAC 3500/MAC 5500 system.
This ensures the network settings are reread from the communications board.
14. Wait at least 10 minutes and verify the MAC 3500/MAC 5500 can
communicate with the MUSE system.
Windows File Server 2000/2003 has an ARP cache that regenerates the MAC
Address/Physical Address table every 10 minutes. When you replace the
communications board, you change the MAC Address/Physical Address. By
waiting at least 10 minutes, you allow the MUSE operating system to
regenerate the MAC Address/Physical Address cache so the MAC 3500/MAC
5500 is able to communicate again with the MUSE system.
15. Perform all necessary functional checkouts.
Table 2. Transmission Information
Field
Value
Use IR for serial line
No
Serial line baud rate
115.2K
Default location
Ethernet (MUSE)
Summary of Contents for MAC 5500
Page 1: ...g MAC 5500 Resting ECG Analysis System Service Manual 2020299 020 Revision E GE Healthcare ...
Page 9: ...Revision E MAC 5500 resting ECG analysis system 1 1 2020299 020 Introduction 1 Introduction ...
Page 10: ...1 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Introduction For your notes ...
Page 55: ...Revision E MAC 5500 resting ECG analysis system 3 1 2020299 020 Installation 3 Installation ...
Page 56: ...3 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Installation For your notes ...
Page 105: ...Revision E MAC 5500 resting ECG analysis system 5 1 2020299 020 Maintenance 5 Maintenance ...
Page 106: ...5 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Maintenance For your notes ...
Page 137: ...Revision E MAC 5500 resting ECG analysis system 6 1 2020299 020 Parts List 6 Parts List ...
Page 138: ...6 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Parts List For your notes ...
Page 175: ...Revision E MAC 5500 resting ECG analysis system Index 1 2020299 020 Index ...
Page 176: ...Index 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 178: ...Index Index 4 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 179: ......