![GE Datex-Ohmeda S/5 Technical Reference Manual Download Page 42](http://html.mh-extra.com/html/ge/datex-ohmeda-s-5/datex-ohmeda-s-5_technical-reference-manual_128582042.webp)
Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
30
Document no
.
M1137263-04
The different mounts to install the Central Unit F-CU5
(1) Wall Mount Hanger for S/5 F-CU5
(2) Power Adapter Mount for F-CU5 Power Unit N-AC
(3) Adapter Plate for S/5 F-CU5
(4) Table Holder for S/5 F-CU5
(5) Holder for S/5 Command Bar
3.4.2 Connecting to mains
Make sure the cables connecting the different units together are connected and secured.
Connect the power cord to the mains power inlet on the N-AC power unit, and to the wall socket.
WARNING
The power cord may only be connected to a three-wire, grounded, hospital
grade receptacle.
3.4.3 Connecting to the Datex-Ohmeda Network
To connect the monitor to the Datex-Ohmeda Network, make sure a Network Board B-UPI4NET
is installed.
Use the Monitor-Network cable to connect the monitor to the network as follows:
1.
Make sure that the power to the Central Unit is turned off.
2.
Connect the RJ-45 network connector to the X10 connector on CPU board, B-CPU6 and
the Identification Plug to the corresponding connector on the B-UPI4NET.
3.
Connect the other RJ-45 connector to the corresponding connector on the wall box.
4.
Turn on the power to the Central Unit. Enter the
Network
service menu:
Monitor Setup
-
Install/Service
(password 16-4-34) -
Service
(password 26-23-8) -
Frame – Network
5.
Make sure that the monitor's network communication has been set according to the used
network software:
Network software S-CNET01 -> DRI Level = 2001
Network software S-CNET02 -> DRI Level = 2001
Network software L-NET03 -> DRI Level = 2003
Network software L-NET05 -> DRI Level = 2005
1
2
4
3
5
Summary of Contents for Datex-Ohmeda S/5
Page 4: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 2 Document no M1137263 04 ...
Page 12: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors vi Document no M1137263 04 ...
Page 28: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 16 Document no M1137263 04 ...
Page 58: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 46 Document no M1137263 04 ...
Page 62: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 50 Document no M1137263 04 ...
Page 86: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 74 Document no M1137263 04 ...
Page 100: ......
Page 102: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1125633 06 ...
Page 108: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1125633 06 ...
Page 126: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 24 Document no M1125633 06 ...
Page 132: ......
Page 136: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1137266 02 ...
Page 216: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 80 Document no M1137266 02 ...
Page 220: ......
Page 258: ......
Page 262: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1125636 05 ...
Page 302: ......
Page 326: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1125637 04 ...
Page 330: ......
Page 332: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137269 02 ...
Page 348: ......
Page 350: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137272 02 ...
Page 384: ......
Page 386: ...Datex Ohmeda S 5TM Anesthesia and Critical Care Monitors ii Document no M1125639 03 ...
Page 400: ......
Page 402: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor ii Document no M1125641 03 ...
Page 416: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor 14 Document no M1125641 03 ...
Page 420: ......
Page 422: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137274 01 ...
Page 428: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1137274 01 ...
Page 431: ...Spare parts 9 Document no M1137274 01 3 2 Central Processor Unit for F CU5 F CPU 02 ...
Page 444: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1137274 01 ...
Page 448: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 26 Document no M1137274 01 ...