Service Menu
37
Document no. M1137266-02
4.7.6 NIBP Pneumatics
How to check Interval 20 mmHg -> 185 mmHg
Select
Start pump
at different combinations of the valves open/closed and push the
ComWheel. The module counts the time it takes for the pressure to go up from 20 mmHg to
185 mmHg and displays it. When all the valves are closed, the pump should be able to pump
the pressure in about 1 to 4 seconds into an adult cuff wrapped around a rigid cylinder or pipe.
The pump does not stop without selecting
Stop Pump
by pushing the ComWheel.
Start Pump/Stop Pump:
A manual control for the
pump. The selection changes to
Stop Pump
when the
pump turns on.
Open Exh1/Close Exh1:
A manual control for the
exhaust valve 1. The selection changes to
Close Exh1
when the valve is opened. This function is not used
with E-PRESTN and E-PSM(P) modules.
Open Exh2/Close Exh2:
A manual control for the
exhaust valve 2. The selection changes to
Close Exh2
when the valve is opened.
Open Zero valve:
A manual control for the zero valve.
This function is not in use with E-PRESTN and E-
PSM(P) modules.
Set Valve:
The opening of the pulse valve is adjusted
between 0 and 255 (0 for fully closed and 255 for fully
open). First push the ComWheel, then turn it to adjust
the value on the screen and finally push to set the
value.
Reset Clock
will zero the time on the display. This
function is not in use with E-PRESTN and E-PSM(P) modules.
Pneumatics Data field
For information on general items
Pressure, Zero, Protect handle, Calibr. prot., +15 V power, AD0
to
AD7
as
well as
Timeouts
etc., see section “
Pump, Exh1 Valve
, and
Exh2 Valve
show their states. NOTE:
Exh1 Valve
has no effect on the module.
Pulse Valve
shows how much the valve is opened (0 to 255) during Valve Setting.
Summary of Contents for Datex-Ohmeda S/5
Page 4: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 2 Document no M1137263 04 ...
Page 12: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors vi Document no M1137263 04 ...
Page 28: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 16 Document no M1137263 04 ...
Page 58: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 46 Document no M1137263 04 ...
Page 62: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 50 Document no M1137263 04 ...
Page 86: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 74 Document no M1137263 04 ...
Page 100: ......
Page 102: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1125633 06 ...
Page 108: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1125633 06 ...
Page 126: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 24 Document no M1125633 06 ...
Page 132: ......
Page 136: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1137266 02 ...
Page 216: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 80 Document no M1137266 02 ...
Page 220: ......
Page 258: ......
Page 262: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1125636 05 ...
Page 302: ......
Page 326: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1125637 04 ...
Page 330: ......
Page 332: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137269 02 ...
Page 348: ......
Page 350: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137272 02 ...
Page 384: ......
Page 386: ...Datex Ohmeda S 5TM Anesthesia and Critical Care Monitors ii Document no M1125639 03 ...
Page 400: ......
Page 402: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor ii Document no M1125641 03 ...
Page 416: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor 14 Document no M1125641 03 ...
Page 420: ......
Page 422: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137274 01 ...
Page 428: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1137274 01 ...
Page 431: ...Spare parts 9 Document no M1137274 01 3 2 Central Processor Unit for F CU5 F CPU 02 ...
Page 444: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1137274 01 ...
Page 448: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 26 Document no M1137274 01 ...