Service Menu
23
Document no. M1137266-02
Balance between leds is adjusted by changing the intensity of red/infrared. Intensity of infrared (
Ired int.
) is in
the range of 40 to 255 and red intensity (
red int.
) is in the range of 40 to 255.
DC gain
shows the gain of DC signal adjusted by the module.
IDC
is the value of infrared signal.
RDC
is the dc value of red signal.
AC gain
is the gain of infrared and red ac signals. AC gain values can be 1 or 0. Value 1 means high ac gain
and 0 means low gain.
Pre gain
is a preamplifier gain for infrared and red signals. Pre gain values can be 1 or 0. Value 1 means
normal operation. Value 0 means that signal levels are very low and extra gain is taken into use.
Temp error
shows the status of the temperature test. No errors found show the status (OFF) and errors found
(ON).
Protect key
shows normally OFF, but turns to ON when the button at the bottom of the module is pressed.
Protect mode
is normally ON.
It turns to OFF, when Protect is switched to OFF for the temperature calibration
in Calibration Menu.
Configuration
shows the chosen module configuration: TP, ST, or STP.
Timeouts
is a cumulative number that indicates how many times the module has not responded to the the
monitor’s inquiry.
Bad checksums
is a cumulative number that indicates how many times communication from the module to
the monitor has broken down.
Bad c-s by mod
is a cumulative number that indicates how many communication errors the module has
detected.
The monitor starts counting these items at power up and resets to zero at power off. The values may also be
reset when a module is attached to the monitor frame and be set to 32769 or continuous counting may be
started when the module is removed from the monitor frame.
The nonzero values do not indicate a failure, but the continuous counting (more than 5 per second) or value
32769 indicates either a serial communication failure or a module not in place. Also failures in other modules
may cause these numbers to rise or be set to 32769.
RAM
indicates the state of the RAM memory.
ROM
indicates whether the checksum at the EPROM is in accordance with the one the software has
calculated.
EEPROM
indicates if the values stored in the permanent memory are valid.
The state is either
OK, Fail
or
?
(module not in place or a communication error).
Summary of Contents for Datex-Ohmeda S/5
Page 4: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 2 Document no M1137263 04 ...
Page 12: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors vi Document no M1137263 04 ...
Page 28: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 16 Document no M1137263 04 ...
Page 58: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 46 Document no M1137263 04 ...
Page 62: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 50 Document no M1137263 04 ...
Page 86: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 74 Document no M1137263 04 ...
Page 100: ......
Page 102: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1125633 06 ...
Page 108: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1125633 06 ...
Page 126: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 24 Document no M1125633 06 ...
Page 132: ......
Page 136: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1137266 02 ...
Page 216: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 80 Document no M1137266 02 ...
Page 220: ......
Page 258: ......
Page 262: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1125636 05 ...
Page 302: ......
Page 326: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1125637 04 ...
Page 330: ......
Page 332: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137269 02 ...
Page 348: ......
Page 350: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137272 02 ...
Page 384: ......
Page 386: ...Datex Ohmeda S 5TM Anesthesia and Critical Care Monitors ii Document no M1125639 03 ...
Page 400: ......
Page 402: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor ii Document no M1125641 03 ...
Page 416: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor 14 Document no M1125641 03 ...
Page 420: ......
Page 422: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137274 01 ...
Page 428: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1137274 01 ...
Page 431: ...Spare parts 9 Document no M1137274 01 3 2 Central Processor Unit for F CU5 F CPU 02 ...
Page 444: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1137274 01 ...
Page 448: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 26 Document no M1137274 01 ...