Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
24
Document no
.
M1137272-02
4.1.2 15” Flat Panel Display, D-FPD15
Please refer also to the display’s Operators Manual.
4.1.3 19” medical grade display
Please refer also to the display’s Operators Manual.
Problem
Cause Treatment
No image on screen
No power.
Check power. Check that both the
patient monitor and the display are ON.
Defective power adapter.
Replace display's power adapter.
Loose or broken video cable.
Tighten or replace video cable.
Defective display controller board.
Replace display controller board.
Defective display.
Replace display.
Picture is unclear
Wrong resolution selected in
AM/CCM menu.
Adjust resolution from the AM/CCM
menu (
Monitor Setup
-
Install/Service
-
Installation
-
Display Setup
)
Jumpers in Display controller board
not positioned correctly.
Check the jumpers. See Installation
section in Part I for details.
Wrong adjustments.
Use auto setup, or position, phase or
clock adjustments in the OSD menu.
Problem
Cause Treatment
No image on screen and no
backlight
No power or loose interface cable.
Check LCD - ON/STBY switch
position and interface cable.
No image on screen, backlight on Cables may be loose. Display
controller board or LCD Interface
board failure.
Check the items. Replace the Display
controller board or change the
display if necessary.
Vertical stripes on right side of
screen continuously
Jumpers or dip switches on Display
controller board not positioned
correctly.
Check the jumpers or dip switches.
See Installation section in Part I for
details.
Picture is unclear.
The picture looks much larger
than the display screen.
Wrong resolution.
Change the resolution to SXGA
Go to monitor setup ->
Install/Service
Æ
Installation
Æ
Display Setup menu select the
resolution. The change is effective
after start-up.
NOTE: The resolution must be set
after every Factory Reset.
Picture is unclear. Colors are not
correct.
Display controller board is not
compatible with the display.
Check that the display controller
board is compatible with the display.
19” medical grade display requires
B-DISPX or B-CPU6.
Summary of Contents for Datex-Ohmeda S/5
Page 4: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 2 Document no M1137263 04 ...
Page 12: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors vi Document no M1137263 04 ...
Page 28: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 16 Document no M1137263 04 ...
Page 58: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 46 Document no M1137263 04 ...
Page 62: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 50 Document no M1137263 04 ...
Page 86: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 74 Document no M1137263 04 ...
Page 100: ......
Page 102: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1125633 06 ...
Page 108: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1125633 06 ...
Page 126: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 24 Document no M1125633 06 ...
Page 132: ......
Page 136: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1137266 02 ...
Page 216: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 80 Document no M1137266 02 ...
Page 220: ......
Page 258: ......
Page 262: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1125636 05 ...
Page 302: ......
Page 326: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1125637 04 ...
Page 330: ......
Page 332: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137269 02 ...
Page 348: ......
Page 350: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137272 02 ...
Page 384: ......
Page 386: ...Datex Ohmeda S 5TM Anesthesia and Critical Care Monitors ii Document no M1125639 03 ...
Page 400: ......
Page 402: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor ii Document no M1125641 03 ...
Page 416: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor 14 Document no M1125641 03 ...
Page 420: ......
Page 422: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137274 01 ...
Page 428: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1137274 01 ...
Page 431: ...Spare parts 9 Document no M1137274 01 3 2 Central Processor Unit for F CU5 F CPU 02 ...
Page 444: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1137274 01 ...
Page 448: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 26 Document no M1137274 01 ...