Service Menu
17
Document no. M1137266-02
4.1.2 Gases
Noise Meas
activates the noise measurement.
Sample gain adj
adjusts the sampling pump gain,
i.e. for adjusting the sample flow
measurement.
Fall time Meas
*activates the fall time
measurement.
Pump ctrl
A manual control for the sampling
pump.
Zero valve ctrl
A manual control for the zero
valve.
Occl valve ctrl*
A manual control for the occlusion
valve.
Record Data
prints out the shown service data
and board information (id. serial
number and software id.) onto the
recorder module.
Service Data
O
2
, CO
2
, N
2
O, AA
%
field shows real-time concentrations.
noise-%
is standard deviation of concentration.
O
2
, CO
2
, N
2
O, AA channels A-E
mV
field: signal is scaled to mV.
Gain
: User gain. It is scaled as (User gain)/(Factory gain).
ID*
Shows the identified agent.
ID unrel.*
The shown value tells how unreliable the identification is. With pure agent the value is
typically < 75.
Sample Flow
is calculated from differential pressure and adjusted by the module.
Zero
value as
measured during initialization when the pump is off.
Gain
: sample flow measurement can be
calibrated by adjusting the gain.
Ambient*
Ambient pressure is measured continuously in E-miniC. In M-miniC it is measured at least
once an hour (during the first running hour during each zeroing process). In M-Gas it is
measured once an hour.
Amb-Work
:
ambient pressure - sampling system internal pressure.
Fall time*
CO
2
and O
2
in ms. For N
2
O and AA same as CO
2
.
CO
2
-O
2
Delay*
In ms. No delay between CO
2
, N
2
O, and agents.
Pump
Can be toggled ON/OFF. PWM output 0-100% is shown. Pump voltage is also shown.
Lamp
The state, PWM control, and current of the lamp are shown.
Fan*
The state of the fan is shown.
Zero and Occl valve
Can be toggled between the measurement state (MEAS) and zeroing/occlusion states
(ZERO/OCCL).
Temp
Temperatures measured by the module from TPX, CPU, and OM.
Time after power on
In minutes after power on.
*)
The function is only in use with Compact Airway Modules.
Summary of Contents for Datex-Ohmeda S/5
Page 4: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 2 Document no M1137263 04 ...
Page 12: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors vi Document no M1137263 04 ...
Page 28: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 16 Document no M1137263 04 ...
Page 58: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 46 Document no M1137263 04 ...
Page 62: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 50 Document no M1137263 04 ...
Page 86: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 74 Document no M1137263 04 ...
Page 100: ......
Page 102: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1125633 06 ...
Page 108: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1125633 06 ...
Page 126: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 24 Document no M1125633 06 ...
Page 132: ......
Page 136: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1137266 02 ...
Page 216: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 80 Document no M1137266 02 ...
Page 220: ......
Page 258: ......
Page 262: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1125636 05 ...
Page 302: ......
Page 326: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1125637 04 ...
Page 330: ......
Page 332: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137269 02 ...
Page 348: ......
Page 350: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137272 02 ...
Page 384: ......
Page 386: ...Datex Ohmeda S 5TM Anesthesia and Critical Care Monitors ii Document no M1125639 03 ...
Page 400: ......
Page 402: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor ii Document no M1125641 03 ...
Page 416: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor 14 Document no M1125641 03 ...
Page 420: ......
Page 422: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137274 01 ...
Page 428: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1137274 01 ...
Page 431: ...Spare parts 9 Document no M1137274 01 3 2 Central Processor Unit for F CU5 F CPU 02 ...
Page 444: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1137274 01 ...
Page 448: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 26 Document no M1137274 01 ...