Planned maintenance check list
11
Document no. M1125633-06
2.2.2 Display
1.
Check that the picture on the screen is adjusted correctly. Readjust the picture with the
adjustment knobs or with the display menu keys, if necessary.
"
2.2.3 Keyboard(s)
1.
Tests with the Command Bar:
−
Press the
Help
key. Turn the ComWheel in both directions and check that the cursor
in the menu moves correspondingly. Select
Normal Screen
and check that the
menu disappears from the screen.
Check the rest of the menu keys by pressing them one by one.
Tests with the ARK Keyboard:
−
Enter the
Keyboard
service menu
−
Check the function of the ComWheel.
−
Press all keys, except
Modify
and
. Check that each key produces a sound
from the loudspeaker, or the ‘Message count’ value in the service menu increases.
−
Press the
Modify
and
keys and check that the corresponding menus appear
on the monitor screen.
Tests with the Remote Controller:
−
Enter the
Keyboard
service menu.
−
Check the function of the Comwheel.
−
Press all keys. Check that each key produces a sound from the loudspeaker, or the
Message count value in the service menu increases.
2.2.4 5-Module Frame, F-CU5(P) / 8-Module Frame, F-CU8
1.
Check that all the fans are running.
2.
Check that the clock on the screen shows correct time.
Readjust the time and date, if necessary.
3.
Enter the
Service Log
service menu.
Check the content of the Service Log for possible problems.
"
Summary of Contents for Datex-Ohmeda S/5
Page 4: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 2 Document no M1137263 04 ...
Page 12: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors vi Document no M1137263 04 ...
Page 28: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 16 Document no M1137263 04 ...
Page 58: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 46 Document no M1137263 04 ...
Page 62: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 50 Document no M1137263 04 ...
Page 86: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 74 Document no M1137263 04 ...
Page 100: ......
Page 102: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1125633 06 ...
Page 108: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1125633 06 ...
Page 126: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 24 Document no M1125633 06 ...
Page 132: ......
Page 136: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1137266 02 ...
Page 216: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 80 Document no M1137266 02 ...
Page 220: ......
Page 258: ......
Page 262: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1125636 05 ...
Page 302: ......
Page 326: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1125637 04 ...
Page 330: ......
Page 332: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137269 02 ...
Page 348: ......
Page 350: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137272 02 ...
Page 384: ......
Page 386: ...Datex Ohmeda S 5TM Anesthesia and Critical Care Monitors ii Document no M1125639 03 ...
Page 400: ......
Page 402: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor ii Document no M1125641 03 ...
Page 416: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor 14 Document no M1125641 03 ...
Page 420: ......
Page 422: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137274 01 ...
Page 428: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1137274 01 ...
Page 431: ...Spare parts 9 Document no M1137274 01 3 2 Central Processor Unit for F CU5 F CPU 02 ...
Page 444: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1137274 01 ...
Page 448: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 26 Document no M1137274 01 ...