![GE Datex-Ohmeda S/5 Technical Reference Manual Download Page 109](http://html.mh-extra.com/html/ge/datex-ohmeda-s-5/datex-ohmeda-s-5_technical-reference-manual_128582109.webp)
Planned maintenance check list
7
Document no. M1125633-06
2
Planned maintenance check list
2.1 Visual inspection/preparation
2.1.1 General
WARNING
Wear a static control wrist strap when handling PC boards. Electrostatic
discharge may damage components on the board.
Make sure that the monitor is switched to standby.
Disconnect the mains power cord. If the monitor is connected to the Datex-Ohmeda
Network, disconnect the Mon-Net cable from CPU board, B-CPU6. If the Memory Module,
E-MEM is connected, remove any memory cards.
1.
Check all the units visually. Check that all parts are intact and that the cables and screws
are connected and tightened properly.
Especially check the following parts:
−
Video displays: the display power cord is locked to the display.
−
F-CU5(P), F-CPU and N-AC: all the screws and the equipotential tap on the N-AC are
tightened properly.
−
F-CU8: the equipotential tap and all the screws are tightened properly.
Check that modules go in smoothly and lock up properly in all module slots.
2.
Replace the batteries, if necessary.
The manufacturer recommendations are:
−
Replace the lead-acid battery in the module frames’ F-CU5(P)/ F-CU8 power supply
unit every 4 years.
−
Replace the SRAM/Timekeeper battery on the CPU board every 8 years.
NOTE: The Factory Reset must be performed if the SRAM/Timekeeper battery, or the
SRAM/Timekeeper chip is replaced.
3.
Clean or replace the fan filters in the 5-Module Frame, F-CU5(P) and Central Processor
Unit, F-CPU or in the 8-Module Frame, F-CU8.
4.
F-CU8: Check that the fuses are of the correct rating.
"
2.1.2 Extension Frame, F-EXT4
1.
If the F-EXT4 contains a fan, clean or replace the fan filter.
"
2.1.3 Compact Airway Module, E-CXX
1.
Detach the module box.
Check that all cables and tubes are connected properly and that there are no loose
objects inside the module.
NOTE: The tubes connected to the Oxygen board pressure transducers should not be pressed
too deep.
Summary of Contents for Datex-Ohmeda S/5
Page 4: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 2 Document no M1137263 04 ...
Page 12: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors vi Document no M1137263 04 ...
Page 28: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 16 Document no M1137263 04 ...
Page 58: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 46 Document no M1137263 04 ...
Page 62: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 50 Document no M1137263 04 ...
Page 86: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 74 Document no M1137263 04 ...
Page 100: ......
Page 102: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1125633 06 ...
Page 108: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1125633 06 ...
Page 126: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 24 Document no M1125633 06 ...
Page 132: ......
Page 136: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1137266 02 ...
Page 216: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 80 Document no M1137266 02 ...
Page 220: ......
Page 258: ......
Page 262: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1125636 05 ...
Page 302: ......
Page 326: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1125637 04 ...
Page 330: ......
Page 332: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137269 02 ...
Page 348: ......
Page 350: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137272 02 ...
Page 384: ......
Page 386: ...Datex Ohmeda S 5TM Anesthesia and Critical Care Monitors ii Document no M1125639 03 ...
Page 400: ......
Page 402: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor ii Document no M1125641 03 ...
Page 416: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor 14 Document no M1125641 03 ...
Page 420: ......
Page 422: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137274 01 ...
Page 428: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1137274 01 ...
Page 431: ...Spare parts 9 Document no M1137274 01 3 2 Central Processor Unit for F CU5 F CPU 02 ...
Page 444: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1137274 01 ...
Page 448: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 26 Document no M1137274 01 ...