Planned maintenance check list
13
Document no. M1125633-06
Patient Spirometry
Compact Airway Modules with the option
11. Enter the
Spirometry
service menu.
Connect a clean spirometry tube and D-lite to the module.
Perform the spirometry leak test.
12. Connect the sampling line. Breathe through the wider side of the D-lite. Check that the
flow waveform moves downwards when you breathe in, and upwards when you breathe
out.
"
General
For all Compact Airway Modules
•
Switch the monitor to standby and reassemble the module.
NOTE: Make sure that all front panel grounding claws are attached and have a good contact
with the module box.
NOTE: When reassembling the module, make sure that the tubes are not pinched between the
module box and internal parts.
13. Insert the Compact Airway Module into the frame. Switch the monitor back on and wait
until the message ‘Calibrating gas sensor’ disappears from the screen.
14. Block the tip of the sampling line with your finger and check that the message ‘Sample
line blocked’ appears on the monitor screen within 30 seconds.
15. Detach the D-fend and check that the message ‘Check D-fend’ appears on the monitor
screen within 60 seconds.
"
2.2.7 Single-width Airway Module, E-miniC
1.
Wait until the message ‘Calibrating gas sensor’ disappears from the screen, then enter
the Gas Unit
General
service menu.
2.
Check that the Time-outs, Bad checksums and Bad c-s by mod values are not increasing
faster than by 5 per second.
3.
Enter the
Gases
service menu: Check that the ‘Ambient’ value displayed corresponds with
the current ambient pressure (±20 mmHg).
4.
Check that the flow measurement offset, i.e. the sample ‘Zero’ value displayed is within
±10 ml/min.
5.
Perform a sampling system leak test.
6.
Check the flow rates. Adjust the sampling flow, if necessary.
7.
Block the tip of the sampling line with your finger and check that the message ‘Sample
line blocked’ appears on the monitor screen within 30 seconds.
Remove the Mini D-fend and check that the message ‘Check D-fend’ appears on the
screen within 30 seconds
.
8.
Perform a gas calibration:
NOTE: For maximum accuracy, a warm-up time of 30 minutes is recommended.
Summary of Contents for Datex-Ohmeda S/5
Page 4: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 2 Document no M1137263 04 ...
Page 12: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors vi Document no M1137263 04 ...
Page 28: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 16 Document no M1137263 04 ...
Page 58: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 46 Document no M1137263 04 ...
Page 62: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 50 Document no M1137263 04 ...
Page 86: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 74 Document no M1137263 04 ...
Page 100: ......
Page 102: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1125633 06 ...
Page 108: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1125633 06 ...
Page 126: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 24 Document no M1125633 06 ...
Page 132: ......
Page 136: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1137266 02 ...
Page 216: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 80 Document no M1137266 02 ...
Page 220: ......
Page 258: ......
Page 262: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1125636 05 ...
Page 302: ......
Page 326: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1125637 04 ...
Page 330: ......
Page 332: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137269 02 ...
Page 348: ......
Page 350: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137272 02 ...
Page 384: ......
Page 386: ...Datex Ohmeda S 5TM Anesthesia and Critical Care Monitors ii Document no M1125639 03 ...
Page 400: ......
Page 402: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor ii Document no M1125641 03 ...
Page 416: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor 14 Document no M1125641 03 ...
Page 420: ......
Page 422: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137274 01 ...
Page 428: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1137274 01 ...
Page 431: ...Spare parts 9 Document no M1137274 01 3 2 Central Processor Unit for F CU5 F CPU 02 ...
Page 444: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1137274 01 ...
Page 448: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 26 Document no M1137274 01 ...