Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
26
Document no
.
M1137266-02
RAM
indicates the state of the RAM memory.
ROM
indicates whether the checksum at the EPROM is in accordance with the one the software has
calculated.
EEPROM
indicates if the values stored in the permanent memory are valid.
The state is either
OK, Fail
or
?
(module not in place or a communication error).
4.4.1 P/PT Calibrations
How to calibrate T3/ T4
The calibrations are possible only when the protection is set
OFF.
The temperature
calibration requires accurate test plugs of value 25 °C and 45 °C.
1.
Select
Calibrate T3 / Calibrate T4.
2.
Insert the test plug 25 °C into the T3 / T4 connector.
3.
Press the ComWheel.
4.
Insert the test plug 45 °C into the T3 / T4 connector.
5.
Press the ComWheel.
How to calibrate P3
The calibration requires a pressure transducer (with an appropriate cable) and a pressure
manometer.
1.
Connect the pressure transducer with the pressure manometer to the P3 connector.
Select
Calibrate P3
. Leave the transducer to room air pressure.
2.
Press the ComWheel to start zeroing.
3.
Supply a pressure of 100 mmHg to 300 mmHg to the transducer. The recommended
pressure is 200 mmHg.
4.
Set the pressure on the display to match the pressure reading on the manometer and
press the ComWheel. A tolerance of ±1 mmHg is allowed.
5.
The message ‘Calibrated’ will appear on the display.
Protection
for the configuration can be set ON and
OFF only when the protect button at the bottom of
the module is pressed.
Set Config:
The module configuration should be
set according to the module type. The setting is
possible only when the protection is set OFF. The
available selections are BP or PT.
The configuration setting should be checked, if the
PT board is replaced.
Calibrate T3 and Calibrate T4:
The functions are
for calibrating the temperature channels T3 and
T4.
Calibrate P3:
The function is for calibrating the
invasive blood pressure channel P3.
Summary of Contents for Datex-Ohmeda S/5
Page 4: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 2 Document no M1137263 04 ...
Page 12: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors vi Document no M1137263 04 ...
Page 28: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 16 Document no M1137263 04 ...
Page 58: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 46 Document no M1137263 04 ...
Page 62: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 50 Document no M1137263 04 ...
Page 86: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 74 Document no M1137263 04 ...
Page 100: ......
Page 102: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1125633 06 ...
Page 108: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1125633 06 ...
Page 126: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 24 Document no M1125633 06 ...
Page 132: ......
Page 136: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1137266 02 ...
Page 216: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 80 Document no M1137266 02 ...
Page 220: ......
Page 258: ......
Page 262: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1125636 05 ...
Page 302: ......
Page 326: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1125637 04 ...
Page 330: ......
Page 332: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137269 02 ...
Page 348: ......
Page 350: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137272 02 ...
Page 384: ......
Page 386: ...Datex Ohmeda S 5TM Anesthesia and Critical Care Monitors ii Document no M1125639 03 ...
Page 400: ......
Page 402: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor ii Document no M1125641 03 ...
Page 416: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor 14 Document no M1125641 03 ...
Page 420: ......
Page 422: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137274 01 ...
Page 428: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1137274 01 ...
Page 431: ...Spare parts 9 Document no M1137274 01 3 2 Central Processor Unit for F CU5 F CPU 02 ...
Page 444: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1137274 01 ...
Page 448: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 26 Document no M1137274 01 ...