Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
56
Document no
.
M1137266-02
Service data in Headbox view
EP Start/Stop
is highlighted when the EP Start/Stop
button is pressed in the headbox or module.
Imp. Check
is highlighted when the Imp. Check
button is pressed in the headbox or module.
Active
shows 1 if the channel is active.
Lead off+
shows the lead off status of all plus
electrodes. 0 means that the lead is on and 1 that
the lead is off.
Lead off-
shows the lead off status of all minus
electrodes. 0 means that the lead is on and 1 that
the lead is off.
Imped+
shows the impedance of plus electrodes of
the last impedance check in all channels.
Imped-
shows the impedance of minus electrodes of
the last impedance check in all channels.
BSR
shows the burst-suppression classification of
EEG waveforms. 0 means artifact, 1 suppression, 2 burst and 3 invalid (the EEG is not classified).
Artefact
shows the artefact status of all channels. 0 means that there are no artefacts on the signal and 1
means that there are artefacts.
Noise
shows 1 if the activity of FEMG is too high for clean EEG signal and that EEG is not probably reliable.
EMG AD
shows the latest AD conversion result of FEMG signal without any filtering or scaling.
Leadset AD
shows the latest AD conversion result of lead set signal without any scaling.
Bipolar
shows the montage type that is active in the module and headbox. 1 means bipolar and 0
referenced.
60 Hz
is 1 if the power frequency filter is set for 60 Hz power frequency.
Imp. Meas.
shows 1 if the impedance measurement is on in the headbox.
Imp. Check Failed
shows 1 if the impedance check has failed.
Mu
EP Meas.
shows 1 if the evoked potential measurement is active in the headbox.
High EP Sampl. Rate
shows 1 if the headbox uses higher sampling rate for 10 ms EP measurement.
EP Epoch Points
shows the number of samples collected for each epoch.
EP Channels
shows the number of channels used in EP measurement.
Hb ROM Error
is 1 if the checksum at the EPROM is not in accordance with the one the software has
calculated.
Summary of Contents for Datex-Ohmeda S/5
Page 4: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 2 Document no M1137263 04 ...
Page 12: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors vi Document no M1137263 04 ...
Page 28: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 16 Document no M1137263 04 ...
Page 58: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 46 Document no M1137263 04 ...
Page 62: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 50 Document no M1137263 04 ...
Page 86: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 74 Document no M1137263 04 ...
Page 100: ......
Page 102: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1125633 06 ...
Page 108: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1125633 06 ...
Page 126: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 24 Document no M1125633 06 ...
Page 132: ......
Page 136: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1137266 02 ...
Page 216: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 80 Document no M1137266 02 ...
Page 220: ......
Page 258: ......
Page 262: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1125636 05 ...
Page 302: ......
Page 326: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1125637 04 ...
Page 330: ......
Page 332: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137269 02 ...
Page 348: ......
Page 350: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137272 02 ...
Page 384: ......
Page 386: ...Datex Ohmeda S 5TM Anesthesia and Critical Care Monitors ii Document no M1125639 03 ...
Page 400: ......
Page 402: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor ii Document no M1125641 03 ...
Page 416: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor 14 Document no M1125641 03 ...
Page 420: ......
Page 422: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137274 01 ...
Page 428: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1137274 01 ...
Page 431: ...Spare parts 9 Document no M1137274 01 3 2 Central Processor Unit for F CU5 F CPU 02 ...
Page 444: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1137274 01 ...
Page 448: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 26 Document no M1137274 01 ...