Service Menu
65
Document no. M1137266-02
4.15 Entropy
Check Sensor
Module measures the impedance of
sensor electrodes. Works the same way as from
Entropy menu or module key.
Module
Service data is changed to the module view.
Sensor
Service data is changed to the sensor view.
Service data in Module view
Measurement
shows whether the Entropy
measurement is ON or OFF. Measurement should
start by itself when the sensor is attached to the
patient.
Last service
shows the date of last maintenance.
Entropy
refers to the respective module key. Is
highlighted when the Entropy key is pressed on the
module. (Opens the Entropy menu.)
Check Sensor
is highlighted when the Check Sensor
key is pressed on the module. Activates the sensor
check.
Mains
shows the currently used mains frequency, 50 or 60 Hz:
BSR Status
shows Burst if the module is not currently detecting suppressed EEG. Shows
Suppression if the module is detecting suppressed EEG periods.
Artefact
shows Off, if there is no high-frequency noise present, On if the module detects noise.
Diathermy
shows On, if the module is detecting diathermy (i.e., electrocautery).
AD clipped
shows On, if the signal is getting clipped at the A/D converter.
Alg. ver. (Mon)
shows the monitor algorithm version number.
Alg. ver. (Mod)
shows the module algorithm version number.
Alg. ver. used
shows the currently used algorithm version.
Timeouts
is a cumulative number that indicates how many times the module has not responded
to the monitor’s inquiry.
Mod Mon Bad Checksum
is a cumulative number that indicates how many times there has been
an error in the message from the module to the monitor.
Mon Mod Bad Checksum
is a cumulative number that indicates how many times there has been
an error in the message from the monitor to the module.
RAM
indicates the state of the RAM memory.
ROM
indicates whether the checksum at the EPROM is in accordance with the one the software
has calculated.
EEPROM
indicates if the values stored in the permanent memory are valid. The states in memory
checks are
OK, Fail
or
?
(module not in place or a communication error).
Summary of Contents for Datex-Ohmeda S/5
Page 4: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 2 Document no M1137263 04 ...
Page 12: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors vi Document no M1137263 04 ...
Page 28: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 16 Document no M1137263 04 ...
Page 58: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 46 Document no M1137263 04 ...
Page 62: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 50 Document no M1137263 04 ...
Page 86: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 74 Document no M1137263 04 ...
Page 100: ......
Page 102: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1125633 06 ...
Page 108: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1125633 06 ...
Page 126: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 24 Document no M1125633 06 ...
Page 132: ......
Page 136: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1137266 02 ...
Page 216: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 80 Document no M1137266 02 ...
Page 220: ......
Page 258: ......
Page 262: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1125636 05 ...
Page 302: ......
Page 326: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1125637 04 ...
Page 330: ......
Page 332: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137269 02 ...
Page 348: ......
Page 350: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137272 02 ...
Page 384: ......
Page 386: ...Datex Ohmeda S 5TM Anesthesia and Critical Care Monitors ii Document no M1125639 03 ...
Page 400: ......
Page 402: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor ii Document no M1125641 03 ...
Page 416: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor 14 Document no M1125641 03 ...
Page 420: ......
Page 422: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137274 01 ...
Page 428: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1137274 01 ...
Page 431: ...Spare parts 9 Document no M1137274 01 3 2 Central Processor Unit for F CU5 F CPU 02 ...
Page 444: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1137274 01 ...
Page 448: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 26 Document no M1137274 01 ...