Open the column installation
wizard
Open the Column Installation trail of the ÄKTA ready Installation wizard.
Action
Step
1
•
UNICORN 5: In the ÄKTA ready System Control module, on the File menu,
click Instant Run and then click Run in the dialog box shown.
•
UNICORN 7: In the ÄKTA ready Method Editor module, on the File menu,
click New Method. Click Method Wizard in the New Method dialog box
and then click OK.
In the wizard, click Column Installation.
2
If you intend to run a column test or a rinsing program, also select the Include
Efficiency Test of Column checkbox and/or Include Rinsing of Column.
(Column testing is discussed in
Section 9.3 Optional column test, on page 135
and rinsing in
Section 9.4 Optional column rinsing, on page 138
.)
3
Select the Print Report Automatically checkbox if you want to automatically
produce a printed report with installation information and test results.
Note:
A printed report can also be produced later. In the Evaluation module, on the
File menu, click Report.
4
Similar to the wizard for installing a new flow kit, the column installation
wizard contains checklists with actions to be performed. Each action is
confirmed by selecting the corresponding Completed checkbox in the wizard.
5
In UNICORN 7 the wizard creates a new method. Edit the column parameters
(see
Edit column parameters in UNICORN 7, on page130
) and save the method
in Method Editor. Then run the method in System Control.
6
128
ÄKTA ready Operating Instructions 28960345 AD
9 Column installation
9.2 Installing a column