Running the column efficiency
test
A column efficiency test is started by clicking Start in the UNICORN Result Name panel
in the Start Protocols dialog box displayed immediately after finishing the installation
wizard. The test is performed using 100 cm/h liquid velocity and normally takes approx-
imately 30 minutes.
If column conditioning is included in the test, it will take longer time than 30
minutes.
Tip:
Similar to a component test, it is possible to enter notes during the test (see
For more information about efficiency testing of columns see the application note Effi-
ciency test of ReadyToProcess columns (see
Section 1.3 Associated documentation, on