Check function of valves
Perform a functional test of all valves in the system by manually switching each valve
on/off in UNICORN. Airtrap valve HV-301 is manually controlled with the AIR VENT button
on the front of the system cabinet.
Check that the flow diagram in UNICORN is updated and that the valve is in its correct
state. In the record table, mark checked function with "OK"/ "Not OK".
Remark
Closed
Open
Valve
XV-001
XV-002
XV-003
XV-004
XV-005
XV-006
XV-021
XV-022
XV-023
XV-031
XV-032
XV-033
XV-051
XV-052
XV-053
XV-054
XV-055
XV-056
HV-301
ÄKTA ready Operating Instructions 28960345 AD
219
14 Reference information
14.6 Checklists