Mount the outlet tubing
Action
Step
XV-051 to XV-
056
1
2
Mount the outlet manifold by insert-
ing each outlet tube into the corre-
sponding outlet pinch valve (valves
XV-051 to XV-056). Do not yet remove
the end caps.
1
Close each valve safety lock by turn-
ing the black handles 180 degrees.
2
Snap the air vent tubing (that runs
from the top of the air trap) onto the
holder (1 in figure) on top of the sys-
tem cabinet, and remove the protec-
tive end cap (2) from the end of the
air vent tubing.
3
Note:
Remember to remove the protective end cap from the air vent tubing. The air
trap will not work properly if the cap is left in place.
Finishing flow kit installation
•
Make sure that the safety locks on all
pinch valves have been properly
closed. There are 19 safety locks in
total.
•
Make sure that the three pressure
flow cells are in place and properly
connected.
•
Set all pinch valves to run position by
turning the Installation switch to
RUN.
•
Check that the end cap on the air
vent tubing of the air trap has been
removed.
ÄKTA ready Operating Instructions 28960345 AD
95
6 Flow kit installation
6.3 Mounting the flow kit