Finish the installation
Action
Step
Turn the Installation switch in
the upper, right corner of the
system cabinet to the COLUMN
INSTALL position. This will open
a free passage from air trap to
column.
1
Let liquid drain from the air trap
down to the column. Release
under-pressure in the air trap
by pressing the AIR VENT button
next to the air trap release valve
(HV-301). If necessary, tap, flick
or squeeze the tubing (arrows
in figure) until it is completely
filled with liquid and no more
bubbles enter the air trap.
2
Turn the Installation switch in
the upper, right corner of the
system cabinet to the RUN posi-
tion.
3
After the column installation wizard has finished, UNICORN displays the start protocol
with information about the installation, and input fields for parameters specifying the
current setup.
The input fields are the same as those for a flow kit installation, as discussed in
installation information, on page 111
, with one exception: There is an additional field for
the column's identification number. Similar to the other identification numbers, this
number can be scanned using a hand scanner, or it can be entered manually.
ÄKTA ready Operating Instructions 28960345 AD
133
9 Column installation
9.2 Installing a column