![GE AKTA ready Operating Instructions Manual Download Page 68](http://html.mh-extra.com/html/ge/akta-ready/akta-ready_operating-instructions-manual_128010068.webp)
CAUTION
Make sure that correct air pressure is always maintained. Too high
or too low air pressure may be hazardous and may cause erroneous
results and leakage.
Instruction
Action
Step
Connect your compressed air supply to the air input connector on the back
of the system, labeled PNEUMATIC AIR SUPPLY. Use the supplied 5 m air
hose with inner diameter 7.5 mm.
1
Switch on the air supply.
2
Check for air leaks.
3
68
ÄKTA ready Operating Instructions 28960345 AD
4 Installation