Gradient flow section types
Two different ÄKTA ready gradient flow sections are available:
•
ÄKTA ready gradient High Flow section, with pump tubing diameter 12 mm and
tubing diameter 9.5 mm.
•
ÄKTA ready gradient Low Flow section, with with pump tubing diameter 8 mm and
tubing diameter 6.4 mm.
Both flow kits use 25 mm TC connectors for connecting to inlet and outlet containers,
pump tubing, and column.
NOTICE
Only use ÄKTA ready flow sections, supplied by GE Healthcare.
Specifications for ÄKTA ready flow kits are found in
ÄKTA ready gradient section specifi-
. Ordering information is in
ÄKTA ready gradient flow section, on
Installation of gradient flow
section
If you have selected the ÄKTA ready gradient checkboxes in the component list (see
) you can follow the wizard to install the ÄKTA ready gradient
flow section. You may also follow the instruction below.
More information about mounting the flow kits can be found in
.
Action
Step
Take the ÄKTA ready pump tubing from
the ÄKTA ready flow kit package. For a de-
scription of the content of the ÄKTA ready
flow kit package refer to
1
100
ÄKTA ready Operating Instructions 28960345 AD
7 ÄKTA ready gradient
7.2 ÄKTA ready gradient flow section