![GE AKTA ready Operating Instructions Manual Download Page 117](http://html.mh-extra.com/html/ge/akta-ready/akta-ready_operating-instructions-manual_128010117.webp)
8.3
Troubleshooting a failed component test
Troubleshooting
Should a component fail the component test, troubleshooting should be done to establish
the cause of failure and allow for correction of the problem. A failing component is
identified in the report or result file by the text TEST FAILED in the component chart.
Troubleshooting is easiest done using Troubleshooting in the installation wizard. The
troubleshooting wizard finishes with an additional component test, identical with the
section previously described.
Open the installation wizard
Action
Step
1
•
UNICORN 5: In the ÄKTA ready System Control module, on the File menu,
click Instant Run and then click Run in the dialog box shown.
•
UNICORN 7: In the ÄKTA ready Method Editor module, on the File menu,
click New Method. Click Method Wizard in the New Method dialog box
and then click OK.
In the wizard, click Troubleshoot Failed Component Test. Select which
components to troubleshoot by selecting one or more of the checkboxes
shown.
2
In UNICORN 7 the wizard creates a new method. Save the method and then
run it in the System Control module.
3