Preparing for recording
2067634-131 Revision E
SEER ™ 1000
47
ATTENTION
DAMAGE IN CONNECTION CABLE AND - Excessive force can damage the
connection and cable.
Never use force to insert the patient cable or the USB cable to the recorder. Do not bend
the patient cable and do not wrap it around the recorder.
If the plug ECG cable or USB download cable is inserted incorrectly, this could cause
serious damage to the machine. The damage caused by improper use are not covered.
Place the stack
ATTENTION
RECORDING OF TIME MIGHT NOT BE ACHIEVED - To achieve the complete recording
period of 7 days, a full battery or a fully charged battery is required. A low quality cell and
without an appointment or an accumulator that is not fully loaded, could not last seven days.
Before each recording the
LED event of buttons
should be green. If it flashes once
Before each recording the
LED event of buttons
should be green. If it flashes once
Before each recording the
LED event of buttons
should be green. If it flashes once
per second, a new battery is needed or a new battery.
ATTENTION
NO BATTERY CHARGER - The recorder is not designed to charge a battery.
Only use the battery when it is fully charged.
Slide the
battery compartment lock
back.
Slide the
battery compartment lock
back.
Slide the
battery compartment lock
back.
Insert the battery or accumulator, as shown below, with the negative end up in the
recorder.