Maintenance
114
SEER ™ 1000
2067634-131 Revision E
1. Before cleaning, the cable pull device.
1. Before cleaning, the cable pull device.
2.
Clean the cables with a cloth moistened with a mild soap solution or alcohol with 70
volumes.
3.
Clean the surfaces with a clean cloth or paper towel.
Disinfect ECG cable
Proceed as follows to disinfect the cable:
1. Before cleaning, the cable pull device.
1. Before cleaning, the cable pull device.
2. Please use the following disinfecting solution, according to the
2. Please use the following disinfecting solution, according to the
APIC guidelines (1996) for the selection and use of disinfectants:
•
Sodium hypochlorite (domestic bleach 5.2%), minimal dilution of 1: 500 (at
least 100 ppm free chlorine) and maximum dilution of 1:10.
•
Any sodium hypochlorite solution that meets all the criteria of the
directive.
3.
Clean the surfaces with a clean cloth or paper towel.
Storing ECG cable
Always store the cables, hanging them vertically in a dry, well-ventilated area. Do not wrap
the cord around the device.
Clean the battery contacts
If the battery voltage, despite using a new battery, or a fully charged battery, appears to be
low, it is advisable to clean the battery contacts. Use a slightly damp cotton swab with alcohol
70 volumes.
Check the ECG cables and connections
Check ECG cable and connections every month, linking them to an ECG simulator.