Problems solution
2067634-131 Revision E
SEER ™ 1000
109
How to inform the service that the recorder is faulty
If the recorder is faulty, this is detected during self-test after turning on. If the self-test during a
fault is detected, the following sequence flashing LED warns that a problem was detected:
•
Both LEDs flash yellow.
•
Both LEDs flash yellow.
•
Follows a 2-second pause.
•
At
The LED of the event keys
flash green.
At
The LED of the event keys
flash green.
At
The LED of the event keys
flash green.
If a defect of this type, please contact the Technical Support GE. Based on the number of times
that the
LED event of keys
flash green, the support service personnel can determine the type of
that the
LED event of keys
flash green, the support service personnel can determine the type of
that the
LED event of keys
flash green, the support service personnel can determine the type of
defect.
The following table shows, which defect is often indicated by a flashing LED.
Error number
(number,
frequency of
blinking LED)
Error
1
The CRC (Cyclic Redundancy Check, self-test firmware) has not bemsucedido. The
firmware must be re-installed.
2
The internal memory is defective or has a problem in the welding points.
3
One or more internal device clocks are not working properly.
4
The exchange of data between the microprocessor and NAND memory is faulty.
5
The NAND memory has too many bad memory blocks.
6
Data exchange with the Bluetooth module is defective.
7
The real time clock had detected a fault. Disconnect and reconnect the recorder. This can
correct the fault.