Page 62
§ 7.4.
ERROR 04 Motor Overrun
04-01
ZL Motor forward
04-02
ZL Motor backward
04-11
ZR Motor forward
04-12
ZR Motor backward
04-21
X Motor forward
04-22
X Motor backward
04-31
Y Motor forward
04-32
Y Motor backward
04-41
Focus X Motor forward
04-42
Focus X Motor backward
04-51
Focus Y Motor forward
04-52
Focus Y Motor backward
Reason1:
Limit switch is damaged
Reason2:
Main PCB is damaged
To diagnose error:
Press [RESET] and [ENT] at the same time. Splicer performs a reset motion
without resetting the motor position. This operation is SECRET, don’t show to customer. Execute
“Input Port Status” command on the secret menu and check the status of error port. Check the
connector terminal using a circuit tester. GND level is ON. By the above procedures, it is possible to
judge if the limit switch or the Main PCB caused the problem.
§ 7.5.
ERROR 05 Close Cover
Reason:
Cover is opened
§ 7.6.
ERROR 06 Cover Opened Error
Reason1:
Connector terminal damaged or not connected
Reason2:
Cover magnet missing or position wrong
Reason3:
Main PCB is damaged
Note
: If cover is opened when [SET] is pressed, ERROR-05 will appear. On the other hand, if the
cover is opened during the image processing, ERROR06 will appear.
To diagnose error:
Execute “Input Port Status” command on the secret menu and check the status of
error port. Check the connector terminal using a circuit tester. GND level is ON. By the above
procedures, it is possible to judge if the limit switch or the Main PCB caused the problem.