Page 63
§ 7.7.
ERROR 07 Motor Trouble
04-01
ZL Motor
04-02
ZR Motor
04-03
X Motor
04-04
Y Motor
04-05
Focus X
04-06
Focus Y
Reason1:
Connector terminal damaged or not connected
Reason2:
Limit switch is damaged
Reason3:
Mechanical adjustment
Reason4:
Motor is damaged
Note:
Error appears when the Motor’s LS dog does not strike the limit switch in the predetermine
duration of time. Standard time is less than 10 seconds for the Z motors, 30 seconds for the XY motors,
and 5 seconds for the focus motors.
To diagnose error:
Execute “Input Port Status” command on the secret menu and check the status of
error port. Check the connector terminal using a circuit tester. If the port and terminal are OK, there is
a possibility that the motor is damaged. Check the motor moving time from forward limit to backward
limit.
§ 7.8.
ERROR 08 Fiber Judge Error
Reason:
Left fiber type and right fiber types are different
Note:
Fiber type is judged only to SM. DS, or MM when using AUTO MODE.
§ 7.9.
ERROR 09 (ARC) Calibration Error
Reason1:
Worn electrodes
Reason2:
Electrode mechanical adjustment