Page 30
§ 4.8.
Adjustment Fiber Clamp Position
1.
Refer to Fig. 4.8.1 for steps 2 ~ 5.
2.
Adjust the Fiber Clamp by positioning the Fiber Clamp over the center of the V-groove.
3.
Attach Fiber clamp with Cap screw (M2×6, T= temporary, WS2).
4.
Remove only one screw and apply screw- locking-glue to the thread of the screw. Tighten Cap
screw (M2×6, T=3, SLG). Repeat process for second Cap screw.
5.
Confirm Fiber Clamp position.
Put it temporary, after adjusting position.
Align the center of them.
Fiber-Clamp
Fiber clamp arm bracket
V-Groove
Two cap screws
(M2×6, T=3, WS2, SLG)
Adjust the both sides.
Fig. 4.8.1