11-2
IT Network
Anti-virus software is not installed on this device.
This device has a single configurable listening port for the purposes of DICOM Echo and Storage
Commitment.
Data flow
DICOM
MWL Server -----------------------> SonoSite X-Porte --------------------------> PACS
Study order Study data
(DICOM MWL) (DICOM Storage)
Please refer to the
SonoSite X-Porte DICOM Conformance Statement
(D10711) for details.
Caution
1
Connection of equipment to an IT network that includes other systems could
result in previously unidentified risks to patients, operators or third parties. Before
connecting the equipment to an uncontrolled IT Network, make sure that all
potential risks resulting from such connections were identified and evaluated, and
suitable countermeasures were put in place. IEC 80001-1:2010 provides
guidance for addressing these risks.
2
When a setting of the IT network to which this device is connected has been
changed, check that the change does not affect this device and take measures if
necessary. Changes to the IT network include:
Changes in network configuration (IP address, router etc.)
Connection of additional items
Disconnection of items
Update of equipment
Upgrade of equipment
Any changes to the IT network could introduce new risks requiring additional
evaluation to be performed as per item 1 above.
Summary of Contents for SONOSITE X-PORTE
Page 1: ...USER GUIDE ...
Page 75: ...Setting up the system 3 39 ...
Page 76: ...3 40 Setting up the system ...
Page 128: ...4 52 Imaging ...
Page 186: ...5 58 Measurements and calculations ...
Page 210: ...6 24 Measurement references ...
Page 234: ...8 16 Cleaning and disinfecting ...
Page 262: ...9 28 Safety ...
Page 334: ...10 72 Acoustic output ...
Page 364: ...B 14 Index printing 5 56 zoom 4 23 ...
Page 365: ......
Page 366: ...P14645 05 P14645 05 ...