10-42
Acoustic output
Table 10-30: Transducer Model: ICTxp, Operating Mode: PW Doppler
Index Label
M.I.
TIS
TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
≤1 A
aprt
>1
Global Maximum Index Value
1.0
-
(a)
-
1.4
(b)
Associated
Acoustic
Parameter
p
r.3
(MPa)
2.01
W
0
(mW)
-
#
19.8
#
min of [W
.3
(z
1
),I
TA.3
(z
1
)] (mW)
-
z
1
(cm)
-
Z
bp
(cm)
-
Z
sp
(cm)
1.66
z@Pll
.3max
(cm)
1.4
d
eq
(Z
sp
)
(cm)
0.20
F
c
(MHz)
4.38
-
#
-
4.36
#
Dim of A
aprt
X (cm)
-
#
-
0.72
#
Y (cm)
-
#
-
0.50
#
Other
Information
PD
(μsec)
1.144
PRF
(Hz)
1302
p
r
@PII
max
(MPa)
2.49
d
eq
@Pll
max
(cm)
0.19
Focal Length
FL
x
(cm)
-
#
-
#
FL
y
(cm)
-
#
-
#
I
PA.3
@MI
max
(W/cm
2
) 263
Operating
Control
Conditions
Control 1: Exam Type
OB
OB
Control 2: Sample Volume Size
1 mm
12 mm
Control 3: Sample Volume
Position
Zone 2
Zone 3
Control 4: PRF
1302 Hz
1562 Hz
(a) This index is not required for this operating mode; value is <1.
(b) This transducer is not intended for transcranial or neonatal cephalic uses.
# No data are reported for this operating condition since the global maximum index value is not reported for the
reason listed. (Reference Global Maximum Index Value line.)
- Data are not applicable for this transducer/mode.
Summary of Contents for SONOSITE X-PORTE
Page 1: ...USER GUIDE ...
Page 75: ...Setting up the system 3 39 ...
Page 76: ...3 40 Setting up the system ...
Page 128: ...4 52 Imaging ...
Page 186: ...5 58 Measurements and calculations ...
Page 210: ...6 24 Measurement references ...
Page 234: ...8 16 Cleaning and disinfecting ...
Page 262: ...9 28 Safety ...
Page 334: ...10 72 Acoustic output ...
Page 364: ...B 14 Index printing 5 56 zoom 4 23 ...
Page 365: ......
Page 366: ...P14645 05 P14645 05 ...