5-26
Measurements and calculations
2
On the
Ao/LA
calculation list, tap the measurement name.
3
Position the calipers by dragging.
4
Save the calculation: Tap
under the measurement name.
To calculate Atrial Volumes (Simpson’s Rule)
1
On a frozen 2D image, tap
Calcs
.
If this control is hidden, tapping
More Controls
displays it.
2
On the calculation list, tap
Atrial Vol
.
3
Do the following for each measurement:
a
On the measurements list, tap the desired view and phase.
b
Drag the caliper to the start position at the annulus.
After lifting your finger from the screen, a pencil icon appears, indicating that the start location is set
and you can begin your trace.
c
Using your finger, trace the atrial cavity (LA or RA).
To make a correction, trace backward on the dotted line.
d
Complete the trace at the other annulus by lifting your finger from the screen.
The trace closes automatically.
e
You can adjust the atrial length by dragging the cursor.
f
To save the calculation, tap
under the measurement name.
Notes
Even after completing the trace, you can still adjust it by dragging the cursor
backward on the dotted line.
The recommended method is to trace from annulus to annulus and allow the
system to automatically close the trace.
Note
Bi-plane volume is displayed when the length of the cavity as measured in the
2- and 4-chamber views differs by less than 5 mm.
Note
To calculate LAVI and/or RAVI, BSA is required.
Summary of Contents for SONOSITE X-PORTE
Page 1: ...USER GUIDE ...
Page 75: ...Setting up the system 3 39 ...
Page 76: ...3 40 Setting up the system ...
Page 128: ...4 52 Imaging ...
Page 186: ...5 58 Measurements and calculations ...
Page 210: ...6 24 Measurement references ...
Page 234: ...8 16 Cleaning and disinfecting ...
Page 262: ...9 28 Safety ...
Page 334: ...10 72 Acoustic output ...
Page 364: ...B 14 Index printing 5 56 zoom 4 23 ...
Page 365: ......
Page 366: ...P14645 05 P14645 05 ...