10-38
Acoustic output
Table 10-26: Transducer Model: HSL25xp (Ophthalmic use), Operating Mode: M Mode
Index Label
M.I.
TIS
TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
≤1
A
aprt
>1
Global Maximum Index Value
0.17
-
0.02
-
0.02
(b)
Associated
Acoustic
Parameter
p
r.3
(MPa)
0.47
W
0
(mW)
-
0.5
0.3
#
min of [W
.3
(z
1
),I
TA.3
(z
1
)] (mW)
-
z
1
(cm)
-
Z
bp
(cm)
-
Z
sp
(cm)
0.95
z@Pll
.3max
(cm)
1.8
d
eq
(Z
sp
)
(cm)
0.13
F
c
(MHz)
7.57
-
7.64
-
7.44
#
Dim of A
aprt
X (cm)
-
0.60
-
0.40
#
Y (cm)
-
0.30
-
0.30
#
Other
Information
PD
(μsec)
0.181
PRF
(Hz)
800
p
r
@PII
max
(MPa)
0.75
d
eq
@Pll
max
(cm)
0.12
Focal Length
FL
x
(cm)
-
1.95
-
#
FL
y
(cm)
-
2.70
-
#
I
PA.3
@MI
max
(W/cm
2
) 12
Operating
Control
Conditions
Control 1: Exam Type
Oph
Oph
Oph
Control 2: Optimization
Gen
Gen
Gen
Control 3: Depth
2.7 cm
4.5cm
3.5cm
(a) This index is not required for this operating mode; value is <1.
(b) This transducer is not intended for transcranial or neonatal cephalic uses.
# No data are reported for this operating condition since the global maximum index value is not reported for the
reason listed. (Reference Global Maximum Index Value line.)
- Data are not applicable for this transducer/mode.
Summary of Contents for SONOSITE X-PORTE
Page 1: ...USER GUIDE ...
Page 75: ...Setting up the system 3 39 ...
Page 76: ...3 40 Setting up the system ...
Page 128: ...4 52 Imaging ...
Page 186: ...5 58 Measurements and calculations ...
Page 210: ...6 24 Measurement references ...
Page 234: ...8 16 Cleaning and disinfecting ...
Page 262: ...9 28 Safety ...
Page 334: ...10 72 Acoustic output ...
Page 364: ...B 14 Index printing 5 56 zoom 4 23 ...
Page 365: ......
Page 366: ...P14645 05 P14645 05 ...