Imaging
4-49
3
(Option) To export images or clips from a closed exam:
a
On the
Review
page, tap
Patient List
.
b
On the
Patient List
page, select a patient record and tap
Thumbnails
.
4
On the
Review
page, select the check box next to each image or clip that you want to export.
5
Tap
Export USB
.
6
To change the file name, tap the
Please enter a file name
box. When the keyboard appears, type the
new file name using upper and lower-case letters and numbers only. Special characters and spaces cannot
be used in a file name.
7
If more than one storage device is installed, tap to select the storage device you want to export to.
8
By default, patient information such as names and IDs are removed from images and clips before
exporting. Patient information can be included during export by selecting the
Include patient
information on images and video clips
check box.
9
Tap
Export
.
ECG
The ECG feature requires an optional FUJIFILM SonoSite ECG module.
ECG is available only on the cardiac exam type.
Note
All images and clips exported to the USB memory stick are stored on the root
level of the drive. This is to facilitate viewing in Image Gallery. Files with the same
name will have the name automatically incremented.
Caution
Patient information maybe a protected class of patient data subject to
country-specific security guidelines. If you choose to include patient
information when exporting images and clips, be sure your information storage
and handling practices comply with country-specific security guidelines.
WARNING
To prevent misdiagnosis, do not use the ECG trace to diagnose cardiac rhythms.
The FUJIFILM SonoSite ECG option is a non-diagnostic feature.
Summary of Contents for SONOSITE X-PORTE
Page 1: ...USER GUIDE ...
Page 75: ...Setting up the system 3 39 ...
Page 76: ...3 40 Setting up the system ...
Page 128: ...4 52 Imaging ...
Page 186: ...5 58 Measurements and calculations ...
Page 210: ...6 24 Measurement references ...
Page 234: ...8 16 Cleaning and disinfecting ...
Page 262: ...9 28 Safety ...
Page 334: ...10 72 Acoustic output ...
Page 364: ...B 14 Index printing 5 56 zoom 4 23 ...
Page 365: ......
Page 366: ...P14645 05 P14645 05 ...