Cleaning and disinfecting
8-5
a
Use
either a premoistened wipe or a soft cloth dampened with cleaner or disinfectant.
Choose a cleaner
from the list of approved cleaners.
b
Remove all gel, debris, and bodily fluids from the transducer.
c
With a new wipe, clean the cable and transducer, starting from the cable, wiping toward the scanhead.
This method helps to avoid cross-contamination.
d
Observe the minimum wet contact time and refer to manufacturer’s instructions. Monitor the transducer
for wet appearance. Re-apply with a new wipe if no longer wet.
Approved cleaners/disinfectants for the transducer:
Product
Compatible transducers
Minimum wet contact time
a
SaniCloth AF3
b
C11xp, C35xp, C60xp, HFL38xp,
HFL50xp, HSL25xp, ICTxp, L25xp,
L38xp, P10xp, P21xp, rP19xp
3 minutes
SaniCloth Plus
C60xp, HFL38xp, HFL50xp, ICTxp
3 minutes
PI-Spray II
C11xp, C60xp, D2xp, HFL38xp,
HFL50xp, HSL25xp, ICTxp, L25xp,
L38xp, P10xp, P21xp, rP19xp
10 minutes
a
For maximum effectiveness, the component being cleaned must remain wet with disinfectant for a minimum
period of time.
b
Qualified for use as an intermediate-level disinfectant for mycobacteria.
Refer to the cleaners and disinfection document available at
www.sonosite.com
for a complete list of the most
current cleaners and disinfectants.
Caution
Do not allow moisture near the electronic components of the connector.
Summary of Contents for SONOSITE X-PORTE
Page 1: ...USER GUIDE ...
Page 75: ...Setting up the system 3 39 ...
Page 76: ...3 40 Setting up the system ...
Page 128: ...4 52 Imaging ...
Page 186: ...5 58 Measurements and calculations ...
Page 210: ...6 24 Measurement references ...
Page 234: ...8 16 Cleaning and disinfecting ...
Page 262: ...9 28 Safety ...
Page 334: ...10 72 Acoustic output ...
Page 364: ...B 14 Index printing 5 56 zoom 4 23 ...
Page 365: ......
Page 366: ...P14645 05 P14645 05 ...