72
t
Click [Start] for the “PS-R” box.
• The alarm beeps intermittently for 10 seconds.
Click [Cancel] to stop the operation.
y
Before the pump starts discharging,
disconnect the PS-R replenisher nozzle
and insert it into the measuring cylinder.
• The replenisher pump starts discharging after
10 second countdown.
• The alarm beeps when discharging is
completed.
u
Insert the PS-R replenisher nozzle into the
PS sub-tank.
i
Place the measuring cylinder on a level
surface, read and record the output
amount.
o
Pour the replenisher solution from the
measuring cylinder into the waste recovery
tank.
!0
Select the “Measured value” box for “PS-
R”, and enter the recorded output value
using the numeric keys or the [
] and [
]
keys.
!1
Repeat steps 5 through 11 for P1W.
The machine automatically measures and adjusts the
pump output for P1-R, P2-RA and P2-RB.
Measurement for P2W is unnecessary since the P2W
pump is not used.
GD1276
PS-R Replenisher Nozzle
Mesearing Cylinder