71
3
SETUP AND MAINTENANCE
q
Prepare a measuring cylinder (200 ml).
w
Remove the sub-tank cover.
e
Open the “6) Processor Adjustment/
Maintenance” menu. (See section 3.1)
r
Select “1 Pump Output Measurement/
Setting”, and press the [START] key.
• The “[461] Pump Output Measurement/Setting”
screen appears.
• The cursor is positioned at “Measured value” for
“P1R”.
3.10
Replenisher Pump Output Measurement
Reduced replenisher pump output may result in improper processing and a drop in
level of solution in the processing tank.
Processing solutions may irritate the eyes and skin. When handling processing solutions, wear
protective goggles, a rubber apron, rubber gloves, rubber boots and other protective gear as
appropriate to prevent any solution from coming in contact with the body. Always wash hands well
after handling processing solutions.
GD1316
Sub-tank Cover