n
The Laboratory - jaw crusher is equipped with a motor protection
switch that is adapted to the mains voltage (according to the type
plate).
The motor protection switch automatically cuts off the current
supply in the event of an overload or a defective motor or cable.
After the malfunction has been rectified, the device can be switched
on again by means of the main switch.
n
The Laboratory - jaw crusher is switched on and off using the main
switch.
– When the main switch is turned to the Start position "I"
®
, the
Laboratory - jaw crusher starts up.
– When the main switch is turned to the Stop position "0"
®
, the
Laboratory - jaw crusher comes to a standstill after approx. 3
seconds.
– The front cover can be opened.
2.8.2 Protection against restart
The Laboratory - jaw crusher comes to a standstill if there is a power
outage during operation: The Jaw Crusher comes to a standstill after
approx. 3 seconds.
When electrical power returns, the Laboratory - jaw crusher does not
start up.
n
The jaw crusher is secured against restarting.
n
When the main switch is turned to the Start position "I": The Labora-
tory - jaw crusher starts up.
Safety information and use