OF-500 OptiFiber Certifying OTDR
Technical Reference Handbook
11-4
Creating Custom Test Limits
You can create up to 10 custom test limits. Each custom
limit can include up to 5 fiber types.
1
Press
F
; then select Edit Custom Test Limit.
2
On the CUSTOM LIMIT LIST screen, do one of the
following:
•
Select an existing custom limit to edit.
•
Press
A
Add Limit to select any existing limit as a
baseline for a new limit.
Note
When you add a new limit, 3 generic fiber types
are placed in the new limit by default.
3
Use the LIMIT TABLE EDITOR screen to see or define
test parameters for the fiber types you will test. Figure
11-2 shows an example of this screen.
•
To see the properties of a factory-installed fiber
type or to edit the properties of a custom fiber
type already used in the limit, use
KNLM
to
highlight a fiber type; then press
t
.
Note
Selecting a factory-installed fiber type lets you see
the fiber’s properties, but you cannot edit them.
•
To add a fiber type to the custom limit press
A
Add Fiber Type; then select a fiber type from the
list or press
A
Create Fiber Type to create a new
type. See the previous section for details on fiber
properties.
•
To remove a fiber type, highlight the fiber; then
press
B
Remove Fiber Type.
4
To add or edit test parameters, use
KNLM
to
highlight a limit cell in the table; then press
t
. Tables
11-2 and 11-3 describe the parameters available for
OTDR and loss/length test limits. It is not necessary to
enter test parameters for tests you will not run.
5
When you are finished entering test parameters on the
LIMIT TABLE EDITOR screen, do one of the following:
•
To save the custom limit using the existing limit
name, press
s
.
•
To edit the name, press
E
Edit Name. Press
s
when
you are finished entering a limit
name; then press
s
again to save the limit.
Summary of Contents for OF-500-01 OptiFiber
Page 4: ......
Page 18: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook xiv...
Page 29: ...Getting Acquainted Powering the Tester 1 9 1 ajt20f eps Figure 1 1 Battery Pack Feature...
Page 31: ...Getting Acquainted Verifying Operation 1 11 1 ajt56f eps Figure 1 2 Removing the Module...
Page 48: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 1 28...
Page 76: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 2 28...
Page 110: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 3 34...
Page 132: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 3 56...
Page 138: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 4 6...
Page 144: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 5 6...
Page 196: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 7 44...
Page 216: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 10 10...
Page 254: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 12 30...
Page 256: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook A 2...
Page 268: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook C 2...