OF-500 OptiFiber Certifying OTDR
Technical Reference Handbook
7-32
W
Caution
If you disconnected the reference test cord from
the source's output, you must set the reference
again to ensure valid measurements.
1
Verify that the settings described in Table 7-1 are
appropriate. Let the tester and source warm up for 5
minutes.
2
Select Loss/Length mode: On the HOME screen, press
A
Change Test; then select Loss/Length.
3
Clean the connectors on the cabling to be tested.
4
Connect the tester to the cabling as shown in Figure
5
Set the source to the correct wavelength and to
continuous wave output.
6
Press
T
.
7
Select a wavelength for the test. Auto is used with
SimpliFiber sources, as described in the next section.
8
To test at a second wavelength, repeat steps 5 through
7.
9
To save the results, press
s
, select or create a fiber
ID; then press
s
again.
If you tested at two wavelengths, both are stored in
the same record.
Using Auto Wavelength Detection with
SimpliFiber Sources (OFTM-5731 and OFTM-5732
Modules)
A Fluke Networks SimpliFiber source can transmit
wavelength information that can be read by OFTM-5731
and OFTM-5732 modules. This feature helps eliminate
wavelength selection errors when testing in Far End Source
mode.
To use auto wavelength detection, do the following:
1
Verify that the SimpliFiber source is set to Auto mode.
The Auto LED should be lit. Press
A
if it is not.
2
Press
Z
to select a wavelength (850 nm or 1300 nm).
For SimpliFiber Pro sources, only one wavelength LED
must be on. The LEDs must not be blinking alternately.
3
Run the Far End Source test, selecting Auto on the
tester's SET SOURCE WAVELENGTH screen.
Summary of Contents for OF-500-01 OptiFiber
Page 4: ......
Page 18: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook xiv...
Page 29: ...Getting Acquainted Powering the Tester 1 9 1 ajt20f eps Figure 1 1 Battery Pack Feature...
Page 31: ...Getting Acquainted Verifying Operation 1 11 1 ajt56f eps Figure 1 2 Removing the Module...
Page 48: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 1 28...
Page 76: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 2 28...
Page 110: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 3 34...
Page 132: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 3 56...
Page 138: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 4 6...
Page 144: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 5 6...
Page 196: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 7 44...
Page 216: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 10 10...
Page 254: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 12 30...
Page 256: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook A 2...
Page 268: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook C 2...