EMC Compliance
This instrument meets the requirements of the EMC Directive 89/336/EEC.
Compliance was demonstrated by meeting the test limits of the following standards:
Emissions
EN61326 (1998) EMC product standard for Electrical Equipment for Measurement,
Control and Laboratory Use. Test limits used were:
a) Radiated:
Class
B
b) Conducted:
Class
B
c) Harmonics:
EN61000-3-2 (2000) Class A
The instrument is Class A by product category.
Immunity
EN61326 (1998) EMC product standard for Electrical Equipment for Measurement,
Control and Laboratory Use. Test methods, limits and performance achieved were:
a) EN61000-4-2
(1995)
Electrostatic Discharge: 4 kV air, 4 kV contact
Performance A.
b) EN61000-4-3
(1997)
Electromagnetic Field: 3 V/m, 80 % AM at 1 kHz
Performance A.
c) EN61000-4-11
(1994)
Voltage Interrupt: 1 cycle, 100 %
Performance A.
d) EN61000-4-4
(1995)
Fast Transient: 1 kV peak (ac line), 0.5 kV peak (signal lines
and RS232/GPIB ports)
Performance A.
e) EN61000-4-5
(1995)
Surge: 0.5 kV (line to line), 1 kV (line to ground)
Performance A.
f) EN61000-4-6
(1996)
Conducted RF: 3 V, 80 % AM at 1kHz (ac line only; signal
connections <3 m not tested)
Performance A.
According to EN61326 the definitions of performance criteria are:
Performance criterion A:
‘During test normal performance within the specification
limits.’
Performance criterion B:
‘During test, temporary degradation, or loss of function or
performance which is self-recovering’.
Performance criterion C:
‘During test, temporary degradation, or loss of function or
performance which requires operator intervention or system
reset occurs.’
Cautions
To ensure continued compliance with the EMC directive the
following precautions should be observed:
a)
connect the generator to other equipment using only high
quality, double-screened cables.
iii
1.888.475.5235
Fluke-Direct
.com
Summary of Contents for DDS Function Generator 271
Page 12: ...271 Users Manual x 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 22: ...271 Users Manual 1 10 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 34: ...271 Users Manual 4 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 46: ...271 Users Manual 5 12 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 52: ...271 Users Manual 6 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 60: ...271 Users Manual 8 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 64: ...271 Users Manual 9 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 76: ...271 Users Manual 12 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 80: ...271 Users Manual 13 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 90: ...271 Users Manual 15 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 110: ...271 Users Manual 16 20 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 112: ...271 Users Manual 2 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 132: ...4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...