271
Users Manual
16-10
Remote Commands
RS232 Remote Command Formats
Serial input to the instrument is buffered in a 256 byte input queue which is filled, under
interrupt, in a manner transparent to all other instrument operations. The instrument will
send
XOFF
when approximately 200 characters are in the queue.
XON
will be sent
when approximately 100 free spaces become available in the queue after
XOFF
was
sent. This queue contains raw (not parsed) data which is taken by the parser as required.
Commands (and queries) are executed in order and the parser will not start a new
command until any previous command or query is complete. In non-addressable RS232
mode responses to commands or queries are sent immediately; there is no output queue.
In addressable mode the response formatter will wait indefinitely if necessary, until the
instrument is addressed to talk and the complete response message has been sent, before
the parser is allowed to start the next command in the input queue.
Commands must be sent as specified in the commands list and must be terminated with
the command terminator code
0AH
(line feed, LF). Commands may be sent in groups
with individual commands separated from each other by the code
3BH
(;). The group
must be terminated with command terminator
0AH
.
Responses from the instrument to the controller are sent as specified in the commands
list. Each response is terminated by
0DH
(carriage return, CR) followed by
0AH
.
<WHITE SPACE>
is defined as character codes
00H
to
20H
inclusive with the
exception of those which are specified as addressable RS232 control codes.
<WHITE SPACE>
is ignored except in command identifiers; for example
*C LS
is
not equivalent to
*CLS
.
The high bit of all characters is ignored.
Commands are not case sensitive.
GPIB Remote Command Formats
GPIB input to the instrument is buffered in a 256 byte input queue which is filled, under
interrupt, in a manner transparent to all other instrument operations. The queue contains
raw (not parsed) data which is taken by the parser as required. Commands (and queries)
are executed in order and the parser will not start a new command until any previous
command or query is complete. There is no output queue which means that the response
formatter will wait, indefinitely if necessary, until the instrument is addressed to talk and
the complete response message has been sent, before the parser is allowed to start the
next command in the input queue.
Commands are sent as
<PROGRAM MESSAGES>
by the controller, each message
consisting of zero or more
<PROGRAM MESSAGE UNIT>
elements separated by
<PROGRAM MESSAGE UNIT SEPARATOR>
elements.
A
<PROGRAM MESSAGE UNIT>
is any of the commands in the remote commands
list.
The
<PROGRAM MESSAGE UNIT SEPARATOR>
is the semicolon character
3BH
(;).
<PROGRAM MESSAGES>
are separated by
<PROGRAM MESSAGE TERMINATOR>
elements which may be any of the following:
1.888.475.5235
Fluke-Direct
.com
Summary of Contents for DDS Function Generator 271
Page 12: ...271 Users Manual x 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 22: ...271 Users Manual 1 10 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 34: ...271 Users Manual 4 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 46: ...271 Users Manual 5 12 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 52: ...271 Users Manual 6 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 60: ...271 Users Manual 8 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 64: ...271 Users Manual 9 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 76: ...271 Users Manual 12 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 80: ...271 Users Manual 13 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 90: ...271 Users Manual 15 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 110: ...271 Users Manual 16 20 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 112: ...271 Users Manual 2 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 132: ...4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...