271
Users Manual
10-2
Staircase
Staircase, or multilevel square waves, are selected by pressing the
STAIR
key; when
STAIR
is selected the lamp beside the key lights. The default staircase is a 4-level
waveform with level changes at 90 ° intervals; to modify or define a new staircase select
the staircase edit menu by pressing the
EDIT
key followed by
STAIR
. When staircase
edit is selected the lamp beside the
STAIR
key flashes to show edit mode; selecting edit
mode always sets staircase on and symmetry to 50 % to permit visual checking of the
waveform.
VALS=ABS AUTO=YES
STEP=00 ACTIVE
LENGTH=0256
LEVEL=+511
The staircase edit menu is shown above. Up to 16 steps can be defined, numbered 00 to
15, with a length and level specified either in absolute terms or as a percentage of full
scale height and cycle length. When the value is set to
ABS
olute in the
VALS
field the
LENGTH
field will accept numbers in the range 0000 to 1024 (the cycle sample length)
and the
LEVEL
field will accept values in the range -512 to +511, i.e. 10-bit resolution
peak-to-peak; -512 and +511 correspond to -10 V and +10 V peaks respectively with the
amplitude on the main menu set to maximum. Note, however, that the actual peak-peak
voltage will be determined by the actual amplitude setting. When the value is set to
%MAX
in the
VALS
field both the length and level fields will accept numbers in the
range 0 to 100 % in 1 % steps.
To edit the staircase, or create a new one, proceed as follows:
Move the cursor to the
STEP
field and use the keyboard or rotary control to select the
first step to be changed; note that the level of the selected step is rapidly dithered during
editing (by an amount equivalent to the least significant bit) to help identify the correct
step on an oscilloscope. Move the cursor to the
LENGTH
field and use the keyboard or
rotary control to enter the new length for that step in the appropriate units; press
ENTER
to enter the value. If the
AUTO
field has been left set at
YES
(the default value) the
cursor will automatically move to the
LEVEL
field; enter a value in the appropriate
units and press
ENTER
again. The cursor will move back to the
LENGTH
field and the
STEP
field will be incremented by 1 ready for the next entry. If
AUTO
has been set to
NO
, the stepping between
LENGTH
and
LEVEL
and the incrementing of the
STEP
must be done manually.
The staircase waveform is made up from steps 00, 01, 02 ... etc., in numeric order, up to
the step whose length brings the total to 1024 or more samples; all these steps, including
any in the sequence that have zero length, will be flagged as
ACTIVE
beside their step
number in the display because changing the
LENGTH
or
LEVEL
of any of them will
affect the waveform. Those steps beyond the last active step will be flagged
INACTIVE
,
even if they have a non-zero length, because changing them will not affect the waveform.
If the length of the last active step takes the total number of samples above 1024 then the
surplus samples are ignored (but the full length is displayed); if the last active sample has
insufficient samples to bring the total to 1024 then the end of the waveform is filled in
with the necessary number of samples at
LEVEL=000
.
Waveform editing forces the symmetry to 50 % to simplify entry; when edit mode is
ended the waveform symmetry will return to that specified on the main menu.
1.888.475.5235
Fluke-Direct
.com
Summary of Contents for DDS Function Generator 271
Page 12: ...271 Users Manual x 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 22: ...271 Users Manual 1 10 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 34: ...271 Users Manual 4 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 46: ...271 Users Manual 5 12 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 52: ...271 Users Manual 6 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 60: ...271 Users Manual 8 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 64: ...271 Users Manual 9 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 76: ...271 Users Manual 12 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 80: ...271 Users Manual 13 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 90: ...271 Users Manual 15 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 110: ...271 Users Manual 16 20 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 112: ...271 Users Manual 2 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 132: ...4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...