Remote Operation
Remote Commands
16
16-15
*TRG
Executes a trigger which will have the same effect as pressing
the
MAN/SYNC
key.
MAN/REMOTE
trigger source must be
selected first. GPIB Group Execute Trigger command (
GET
)
will perform the same function as
*TRG
.
System Commands
BEEPMODE <cpd>
Set beep mode to
<ON>
,
<OFF>
,
<WARN>
or
<ERROR>
BEEP
Sound one beep.
*RCL <nrf>
Recalls the instrument set up contained in store number
<nrf>
. Valid store numbers are
0
to
9
. Recalling store
0
sets
all parameters to the factory default settings.
*RST
Resets the instrument parameters to their factory default
values.
*SAV <nrf>
Saves the complete instrument set up in the store number
<nrf>
. Valid store numbers are
1
to
9
.
Status Commands
*LRN?
Returns the complete set up of the instrument as a
hexadecimal character data block approximately 842 bytes
long. To re-install the set up the block should be returned to
the instrument exactly as it is received. The syntax of the
response is
LRN <character data><rmt>
.The settings
in the instrument are not affected by execution of the
*LRN?
command.
LRN <character
data>
Install data for a previous
*LRN?
command.
EER?
Query and clear execution error number register. The
response format is <
nr1><rmt>
.
QER?
Query and clear query error number register. The response
format is <
nr1><rmt>
*CLS
Clear status. Clears the Standard Event Status Register, Query
Error Register and Execution Error Register. This indirectly
clears the Status Byte Register.
*ESE <nrf>
Set the Standard Event Status Enable Register to the value of
<nrf>
.
*ESE?
Returns the value in the Standard Event Status Enable
Register in
<nr1>
numeric format. The syntax of the
response is
<nr1><rmt>
*ESR?
Returns the value in the Standard Event Status Register in
<nr1
> numeric format. The register is then cleared. The
syntax of the response is
<nr1><rmt>
*IST?
Returns
ist
local message as defined by IEEE Std. 488.2.
The syntax of the response is
0<rmt>
, if the local message
false or
1<rmt>
, if the local message is true.
*OPC
Sets the Operation Complete bit (bit 0) in the Standard Event
Status Register. This will happen immediately the command
is executed because of the sequential nature of all operations.
1.888.475.5235
Fluke-Direct
.com
Summary of Contents for DDS Function Generator 271
Page 12: ...271 Users Manual x 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 22: ...271 Users Manual 1 10 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 34: ...271 Users Manual 4 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 46: ...271 Users Manual 5 12 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 52: ...271 Users Manual 6 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 60: ...271 Users Manual 8 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 64: ...271 Users Manual 9 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 76: ...271 Users Manual 12 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 80: ...271 Users Manual 13 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 90: ...271 Users Manual 15 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 110: ...271 Users Manual 16 20 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 112: ...271 Users Manual 2 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 132: ...4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...