Users Manual
16-2
Remote Operation
The following sections detail the operation of the instrument via both the GPIB and the
RS232 interfaces. Where operation is identical no distinction is made between the two.
Where differences occur these are detailed in the appropriate sections. It is therefore only
necessary to read the general sections and those sections specific to the interface of
interest.
Address and Baud Rate Selection
For successful operation, each instrument connected to the GPIB or addressable RS232
system must be assigned a unique address and, in the case of addressable RS232, all must
be set to the same Baud rate.
The instruments remote address for operation on both the RS232 and GPIB interfaces is
set via the menu accessed by pressing the
REMOTE
button.
REMOTE=RS232
ADDRESS=05
BAUD RATE=9600
With the edit cursor in the
REMOTE
field, the selected interface can be toggled between
RS232
and
GPIB
with alternate presses of the
DIGIT
keys, or by using the rotary
control.
The address is selected with the edit cursor in the
ADDRESS
field, using the
DIGIT
keys or rotary control.
Lastly the baud rate is selected with the edit cursor in the
BAUD RATE
field, again
using the
DIGIT
keys or rotary control.
When operating on the GPIB all device operations are performed through a single
primary address; no secondary addressing is used.
Note
GPIB address 31 is not allowed by the IEEE 488 standards but it is
possible to select it as an RS232 address.
Remote/Local Operation
At power-on the instrument will be in the local state with the
REM
ote lamp off. In this
state all keyboard operations are possible. When the instrument is addressed to listen and
a command is received the remote state will be entered and the
REM
ote lamp will be
turned on. In this state the keyboard is locked out and remote commands only will be
processed. The instrument may be returned to the local state by pressing the
EDIT
key
(which doubles as the
LOCAL
key); however, the effect of this action will remain only
until the instrument is addressed again or receives another character from the interface,
when the remote state will once again be entered.
1.888.475.5235
Fluke-Direct
.com
Summary of Contents for DDS Function Generator 271
Page 12: ...271 Users Manual x 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 22: ...271 Users Manual 1 10 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 34: ...271 Users Manual 4 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 46: ...271 Users Manual 5 12 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 52: ...271 Users Manual 6 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 60: ...271 Users Manual 8 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 64: ...271 Users Manual 9 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 76: ...271 Users Manual 12 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 80: ...271 Users Manual 13 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 90: ...271 Users Manual 15 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 110: ...271 Users Manual 16 20 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 112: ...271 Users Manual 2 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 132: ...4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...