Calibration
Calibration Procedure
14
14-3
Using the Password to Access Calibration or Change the Password
With the password set, pressing
EDIT
following by
CAL
will now display the following
screen:
ENTER PASSWORD
----
When the correct password has been entered from the keyboard the display changes to
the opening screen of the calibration routine itself and calibration can proceed as
described below under
Calibration Routine
. If an incorrect password is entered the
message
INCORRECT PASSWORD!
is shown for two seconds before the display
reverts to the main menu.
With the opening screen of the calibration routine displayed after correctly entering the
password, the password can be changed as follows:
Press
EDIT
again; the display changes to
ENTER PASSWORD
. Enter the current
password again; the display changes to
ENTER NEW PASSWORD
. Enter the new
password as described above.
The password is held in EEPROM and will not be lost when the memory battery back-up
is lost. In the event of the password being forgotten, contact the manufacturer for help in
resetting the instrument.
Calibration Routine
The
CAL
ibration procedure proper is accessed by pressing the
EDIT
key followed by
CAL
, the numeric
6
key. At each step the display changes to prompt the user to adjust
the rotary control or the
FIELD
or
DIGIT
keys, until the reading on the specified
instrument is at the value given.
Two keys and the rotary control are used to make adjustments: the
FIELD
keys provide
very coarse adjustment, the
DIGIT
keys coarse adjustment and the rotary control fine
adjustment.
Pressing
ENTER
increments the procedure to the next step; pressing
CE
decrements to
the previous step.
Alternatively, pressing
ESCAPE
exits to the last
CAL
display, at which the user can
choose either to keep the new calibration values by pressing
ENTER
, to return to the old
values by pressing
ESCAPE
or to restart the calibration procedure by pressing
CE
.
The first two displays,
CAL 00
and
CAL 01
, specify the connections and adjustment
method. The subsequent displays,
CAL 02
to
CAL 20
, permit all adjustable
parameters to be calibrated.
The full procedure is as follows:
CAL 02
Zero dc offset.
Adjust for 0 V ±5 mV
CAL 03
dc ve full scale.
Adjust for 10 V ±20 mV
CAL 04
dc offset
−
ve full scale.
Check for
−
10 V ±20 mV
CAL 05
Multiplier zero offset.
Adjust for minimum
1.888.475.5235
Fluke-Direct
.com
Summary of Contents for DDS Function Generator 271
Page 12: ...271 Users Manual x 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 22: ...271 Users Manual 1 10 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 34: ...271 Users Manual 4 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 46: ...271 Users Manual 5 12 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 52: ...271 Users Manual 6 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 60: ...271 Users Manual 8 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 64: ...271 Users Manual 9 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 76: ...271 Users Manual 12 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 80: ...271 Users Manual 13 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 90: ...271 Users Manual 15 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 110: ...271 Users Manual 16 20 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 112: ...271 Users Manual 2 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 132: ...4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...