271
Users Manual
16-14
Staircase and Arbitrary Waveforms
STAIR
Set staircase function
SETSTAIR
<nrf>,<nrf>
Define a new staircase function. Up to 16 pairs of length and
level may be specified; valid length range 0000 to 1024, valid
level
−
512 to +511.
ARB
Set arbitrary function
SETARB
<nrf>,<nrf>
Define a new arbitrary function. 1024 values must be
specified to set the waveform, each one a level in the range -
512 to +511.
ARBSAV
<nrf>,<cpd>
Save arbitrary waveform to store
<nrf>
with name
<cpd>
.
The maximum length of the name is 16 characters.
N.B. If it is required to retain a waveform sent by a
SETARB
command,
ARBSAV
must be used immediately after
SETARB
. If this is not done, any other ARB operation except
ARB
will destroy the data. The waveform data will also be
lost at power down unless it is saved first.
ARBRCL <nrf>
Recall arbitrary waveform from store
<nrf>
ARB?
Query the selected arbitrary waveform; responds
SETARB
<1024 nr1><rmt>
Waveform Generation Options
SQRWAVGEN <cpd>
Set square wave generation mode to
<AUTO>
,
<HF>
or
<LF>
FILTER <cpd>
Set filter mode to
<AUTO>
,
<ON>
or
<OFF>
AUX <cpd>
Set AUX output mode to
<AUTO>
,
<HF>
or
<LF>
SWPTRGOUT <cpd>
Set sweep/tgen output bnc mode to
<AUTO>
,
<SWEEP>
or
<TGEN>
Hop Commands
HOP <cpd>,<nrf>
Set hop status to
<RUN>
or
<OFF>
with last step set to
<nrf>
.
SETHOP
<nrf>,<nrf>,
<nrf>,<nrf>,
<cpd>,<nrf>
Data for one step in the sequence :
<step>
,
<time>
,
<freq>
,
<level>
,
<func>
,
<offset>
.
<step>
is the step number to be defined.
<time>
is the time in seconds to remain in this step. If set to
0
,
MANUAL
will be selected. If set to
le-3 EXTERNAL
will be selected.
<freq>
is the main generator frequency in Hz.
<level>
is the output level expressed in V p-p into an open
circuit.
<func>
is any of
<SINE>
,
<SQUARE>
,
<TRIAN>
,
<POSPUL>
,
<NEGPUL>
,
<POSRAMP>
,
<NEGRAMP>
,
<STAIR>
or
<ARB>
.
<offset>
is the dc offset in volts.
1.888.475.5235
Fluke-Direct
.com
Summary of Contents for DDS Function Generator 271
Page 12: ...271 Users Manual x 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 22: ...271 Users Manual 1 10 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 34: ...271 Users Manual 4 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 46: ...271 Users Manual 5 12 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 52: ...271 Users Manual 6 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 60: ...271 Users Manual 8 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 64: ...271 Users Manual 9 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 76: ...271 Users Manual 12 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 80: ...271 Users Manual 13 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 90: ...271 Users Manual 15 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 110: ...271 Users Manual 16 20 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 112: ...271 Users Manual 2 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 132: ...4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...