Remote Operation
Remote Operation
16
Status Model
sha0020f.emf
Status model
Power on Settings
The following instrument status values are set at power on:
Status Byte Register
= 0
* Service Request Enable Register
= 0
Standard Event Status Register
= 128 (pon bit set)
* Standard Event Status Enable Register
= 0
Execution Error Register
= 0
Query Error Register
= 0
* Parallel Poll Enable Register
= 0
*Registers marked with an asterisk are specific to the GPIB section of the instrument and
are of limited use in an RS232 environment.
The instrument will be in local state with the keyboard active.
The instrument parameters at power on are determined by the setting of the
POWER UP
field on the
SYS
tem menu, as described in the
System Operations
chapter. If
POWER UP=POWER DOWN
or
POWER UP=RECALL nn
has been set and a defined
state is required by the controller at start up then the command
*RST
should be used to
load the system defaults.
If for any reason an error is detected at power up in the non-volatile ram a warning will
be issued and all settings will be returned to their default states as for a
RST
command.
16-9
1.888.475.5235
Fluke-Direct
.com
Summary of Contents for DDS Function Generator 271
Page 12: ...271 Users Manual x 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 22: ...271 Users Manual 1 10 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 34: ...271 Users Manual 4 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 46: ...271 Users Manual 5 12 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 52: ...271 Users Manual 6 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 60: ...271 Users Manual 8 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 64: ...271 Users Manual 9 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 76: ...271 Users Manual 12 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 80: ...271 Users Manual 13 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 90: ...271 Users Manual 15 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 110: ...271 Users Manual 16 20 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 112: ...271 Users Manual 2 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 132: ...4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...